SP9 ServiceManual.docx 06/02/2015
Page 8 of 42
Operating SP9 Undercounte r Unit for the Firs t T ime:
Check that all installation procedures have been carried out.
Ensure water is connected to unit, valve is turned on and power cord is connected and
plugged in to wall socket.
The “power on” light will glow green and the machine will fill to a safe level, above the
elements, before heating. The display will show the current water temperature and the
status “Filling…”
The “Ready/Status” light will cycle two red flashes while the machine is filling to the safe
level.
After this amount of water has heated to about 94ºC, the boiler will draw more water in until
the temperature drops by 1 or 2 degrees. The boiler will then heat again. This heat fill cycle
continues until the boiler is full.
Ensure the Undercounter unit is set to SP9 mode for use with the SP9 brewer. See Menu
Navigation section in this manual
While
filling, the “Ready/Status” light will remain blank.
The “Ready/Status” light will glow green when the machine is both full and up to normal
operating temperature. Allow approximately 15 minutes.
The boiler is now ready for use
– the display will show the Water Temperature and the
status “READY”.
5. SETUP AND CALIBRATION:
Operating the Head Unit for the first time:
To help prevent heat loss between the Undercounter unit and the Head unit the silicone hose can
be insulated with hose insulation. Simply cut the insulation to the correct length required.
Once the Head Unit installation procedure has been carried out, place a 1 litre jug under the
dispense head. Set the dispense volume to 750ml and press the start switch twice. This will
dispense 750ml in one delivery, without pausing. Measure the dispensed volume to confirm factory
calibration (750ml
+
10ml). If the volume is not satisfactory, proceed to the calibration procedure
below to calibrate the Head Unit. The calibration should only be conducted by a trained service
engineer. See guidelines below.
Dispense Volume Calibration
1. Once all installation procedures are carried out the Head Unit is ready to be calibrated.
2. Turn volume dial to Max. Set Brew Time to 2 minutes.
3. Place large container (>1 litre) of known weight under spray head.
4. Initiate brew.
5. Weigh & record brew output.
6. The output should be 750g, plus or minus 10g.
7. If output is within 10g of 750g calibration is complete.
8. If not, enter calibration mode by holding down toggle switch for min 3 sec.
9. A long beep will sound to confirm the unit is in calibration mode.
10. Enter the volume recorded via the brew switch using the following example:
If 724g was measured
The toggle switch is used to input each digit successively
Summary of Contents for 1000830
Page 7: ...SP9 ServiceManual docx 06 02 2015 Page 7 of 42 Figure 3 Figure 4...
Page 11: ...SP9 ServiceManual docx 06 02 2015 Page 11 of 42 CONTROLS Figure 6 BREWING...
Page 20: ...SP9 ServiceManual docx 06 02 2015 Page 20 of 42...
Page 21: ...SP9 ServiceManual docx 06 02 2015 Page 21 of 42...
Page 24: ...SP9 ServiceManual docx 06 02 2015 Page 24 of 42 Internal Component Map Figure 13...
Page 31: ...SP9 ServiceManual docx 06 02 2015 Page 31 of 42 15 ELECTRICAL SCHEMATICS...
Page 32: ...SP9 ServiceManual docx 06 02 2015 Page 32 of 42...
Page 33: ...SP9 ServiceManual docx 06 02 2015 Page 33 of 42 16 SPARE PARTS SCHEMATICS...
Page 35: ...SP9 ServiceManual docx 06 02 2015 Page 35 of 42...
Page 36: ...SP9 ServiceManual docx 06 02 2015 Page 36 of 42...
Page 37: ...SP9 ServiceManual docx 06 02 2015 Page 37 of 42 SP9 Head Unit Spare Parts...
Page 39: ...SP9 ServiceManual docx 06 02 2015 Page 39 of 42 18 DIMENSIONED DRAWINGS...
Page 40: ...SP9 ServiceManual docx 06 02 2015 Page 40 of 42...
Page 41: ...SP9 ServiceManual docx 06 02 2015 Page 41 of 42 19 CUT OUT TEMPLATE...
Page 42: ...SP9 ServiceManual docx 06 02 2015 Page 42 of 42 Installation Cut Out Template...