Cleaning and disinfection
Disassembly
24
GA 5750.4172 EN 14
6
6.2
Disassembly
CAUTION!
Improper cleaning can cause property damage!
Disassemble inhalation mask, nebulizer housing, profile gasket and medicine cup
for reconditioning. Separate nozzle, suction sleeve and piece of tube to avoid that
the parts stick together during reconditioning.
1
3
4
7
2
6
5
8
Fig. 13: Disassembly
Disassembly
Remove inhalation mask (1) and adapter
(2).
Unscrew nebulizer housing (3) and
medicine cup (4).
Remove profile gasket (5), turn off nozzle
(6), separate suction sleeve (7) and piece
of tube (8).
6.3
Cleaning
6.3.1
General
NOTE
In the event of very dirty product surfaces, carry out an additional disinfection
procedure before cleaning the product.
NOTE
Use only all-purpose cleaners which are slightly alkaline (soap solution) and
contain tensides and phosphates as the active cleaning agents.
In the event of heavily contaminated surfaces, use concentrated multi-purpose
detergent.
CAUTION!
Improper cleaning can cause property damage!
Residues of physiological saline solutions (e. g. sodium chloride) can attack the
surfaces of the product.
Remove residues of physiological saline solutions with a cloth dipped in clean
water. Then dry the product with a dry, lint free cloth.