Mantracourt Electronics Limited
PSDS User Manual
56
Range 1
–
6 Tabs
A tab exists for each of the six calibration ranges. Click on a tab to view, edit or start a calibration.
The common controls will be described here but view the individual sections for specifics regarding each
calibration type.
It is important to select Sensitivity before starting a calibration. If sensitivity is changed after
calibration this may result in incorrect measurement.
Item
Description
Name
edit field
Shows the name assigned to this calibration range. Click to enter a new name.
Mode
label
Indicates the type of calibration used internally. By default, the calibration modes are
calibrated for 6 wire connections to eliminate the effect of cable length on the calibration.
See connection section of the manual where it explains how to wire up for a 4 wire connection
to the strain bridge sensor.
Sensitivity
Drop down
Choose a sensitivity range suitable for the connected strain gauge. Choose the lowest range
that exceeds the sensitivity of the connected bridge:
±7.5 mV/V
±15 mV/V
±30 mV/V
±60 mV/V
±120 mV/V
±240 mV/V
±480 mV/V