1.0 Specifications
Measurement method
Ratio determination using a primary nephelometric light scatter signal
(90°) to the transmitted light scatter signal.
Reading units
NTU/ EBC
Lamp Source
Tungsten Lamp
Detector
Silicon Photocell
Measuring Range
0 to1000 NTU/ 0 to 250 EBC
Resolution
0,01 on lowest range
Accuracy
± 2% of reading: 0 to 500 NTU / ± 3% 500 to 1000 NTU
Repeatability
± 1% of reading or 0,01 NTU
Automatic Reading
With user-defined intervals 0 to 250 seconds
Maximum uncertainty
± 2% of full scale
Display
LCD 2 lines / 16 characters
Response Time
Programmable 6 to 41 sec
Data Logger
Up to 1000 data
Auto ShutOff
Programmable from 1 to 60 min
Fast Cal function
Quick Calibration for single point
Software Functions
Signal averaging, "Fast Settling", results freezing, analyst and sample
identification, calibration status, verification and calibration reminder,
calibration history, password
Sample Required
15 ml vial with lid
Sample vials
Round borosilicate glass vial with screw and caps (Φ = 24,5 mm)
Power Supply
4 AA batteries or power supply cable
Indicator
Low battery indicator / battery exchange
Serial Output
USB
Enclosure rating
IP-67, waterproof
Storage Conditions
0 to 40°C ( instrument only)
Dimensions
114 x 198 x 83 mm
Parts and Accessories - Included.
3 vials of 24,5 mm
Batteries AA e power supply cable
Instruction Manual and Carrying Case
Parts and Accessories - Optional.
Backlight Display
Silicone oil
Cleaning Cloth