5.3.8 Schedule Cal.
Access this funcntion to set up the time (Days/hours) before calibration is requested.
5.3.8.1 F.Scale
Choose the number of days and hours before the calibration warning graph will appear
on the display using ▲ , ▼ and
MENU/READ
(send the cursor to the right/left), press and
hold
SAVE
for 3 seconds to store the data.
NOTE: When the calibration schedule is programmded, a graph will be displayed in
the upper right corner during measurements, when the calibration date arrives, a
bar will appear in the graph and a Calibration warning will be displayed.
5.3.9 Customize
Access this funcntion to insert an Nane for the unit, this name will show up during instrument
start up and will be avaliable wen downloading the data.
-
Select tha name you want to give to the unit Using ▲ to scroll to letters , ▼ to scroll to
numbers (both can be used to go up or down) and
MENU/READ
(send the cursor to the
right/left), press and hold
SAVE
for 3 seconds to store the data..
5.3.10 Pat Number
Access this funcntion to insert an Nunber for the unit, this number will be avaliable wen
downloading the data.
-Select tha number you want to give
to the unit Using ▲ to scroll to letters , ▼ to scroll to
numbers (both can be used to go up or down) and
MENU/READ
(send the cursor to the
right/left), press and hold
SAVE
for 3 seconds to store the data.Language
Access this funcntion change the language of the display
Using ▲ or ▼ select from US (English) , ES (Spanish) or BR (Portuguese) when done, press
and hold
SAVE
for 3 seconds to store the data and
ESC
to return to the previous menu.
5.3.10 Instrument ID
Use ▲ , ▼ and
MENU/READ
(send the cursor to the right/left) to set an ID for the unit, press
and hold
SAVE
for 3 seconds to store the data.