Unplug  the  Airflo  from  the  battery  charger.  

Clean the external surfaces using a damp cloth 

and allow to dry thoroughly before reconnecting 

to the battery charger. Do not allow moisture to 

enter any part of the compressor.
Do not use solvent-based products.

Storage

When not in use, store in a dry place together 

with these instructions. To maintain the batteries, 

when  being  stored  for  long  periods,  put  the 

Airflo on charge overnight every 6 months.

 

Maintenance - fig.  

11

At  approximately  12  month  intervals,  inspect  the 

air filter located in the base of the Airflo (

A

), and if 

necessary, fit a new filter (a kit of replacement filters 

is  available  through  your  supplier  -  part  number: 

CA0233).
Battery replacement:

Turn the Airflo upside down and remove the four 

screws (

B

) followed by the two battery supports 

and rubber spacers (

C

). Turn the  Airflo the right 

way up to remove the batteries.
Fit  the  replacement  batteries  so  that  their 

contacts  are  towards  the  center  line  of  the 

battery  compartment. 

NOTE:

  Be  careful  not 

to displace the battery compartment terminals 

when changing the batteries.

IMPORTANT:  Always  replace  both  batteries  and 

use the same type.
The Airflo contains no other serviceable parts. If a 

fault should develop, please contact your supplier 

for further advice.
For  details  of  Mangar  International’s  full 

range  of  Customer  Services  please  contact: 

[email protected]

 or contact your 

local supplier.

B

A

C

11

Summary of Contents for airflo

Page 1: ...user instructions and warranty airflo battery version freedom through lightweight solutions...

Page 2: ...in the presence of flammable anaesthetic mixtures with air oxygen or nitrous oxide PLEASE BE SURE THAT YOU HAVE READ AND UNDERSTOOD THESE INSTRUCTIONS AND THOSE PROVIDED WITH ANY OTHER MANGAR INTERNA...

Page 3: ...re that the battery charger is not connected to the Airflo and then press the battery indicator button A to activate the indicator lights B If the batteries are fully charged all four indicator lights...

Page 4: ...ed If convenient the Airflo may remain continuously on charge when not in use as the batteries cannot be overcharged using this charger WARNING DO NOT recharge the batteries in a bathroom or in a damp...

Page 5: ...AMEL Press the switch to change between Normal and High air flow Normal High WARNING DO NOT operate the Normal High air flow switch while the compressor is running Different ways to operate the Airflo...

Page 6: ...t a battery voltage check each time air is required and if the voltage becomes too low the Auto function will be cancelled automatically Locking the built in buttons fig 7 If desired the built in butt...

Page 7: ...Auto Control Button Move the lever C to up to raise Move the lever to down to lower Move the lever to neutral stop to stop at any height CAMEL and ELK Hand Controls fig 10 Connect the four hoses A fr...

Page 8: ...a new filter a kit of replacement filters is available through your supplier part number CA0233 Battery replacement Turn the Airflo upside down and remove the four screws B followed by the two battery...

Page 9: ...ir flow switch is in the Normal position Auto light flashing This happens when the compressor s internal temperature cut out has tripped Leave the Airflo for 15 minutes to cool The top two lights on t...

Page 10: ...nt AC Fuse with rating Max Min operating environment temperature Auto select button Up button Down button Do Not connect to mains power in a bathroom or outdoors Class II Double Insulated equipment Da...

Page 11: ...or repair by unqualified persons The following statements do not affect the consumer s statutory rights This product is sold on the understanding that in the event of any defect of manufacture or mat...

Reviews: