Mangar airflo User Instructions And Warranty Download Page 6

airflo (battery version)

6

ME0082 issue 1

7

A

A

B

C

D

E

B

8

‘Auto’ Control Button - fig.  

7

The  ‘Auto’  function  is  designed  for  use  with  the 

range  of  Mangar  International  pneumatic  hand 

controls:-

Lever Hand Control - fig. 

9

CAMEL Hand Control - fig. 

10

ELK Hand Control - fig. 

10

To switch on, press and hold the ‘Auto’ button until 

the light (

A

) comes on. The light will stay on while 

the ‘Auto’ function is selected.

Pressing  any  of  the  built-in  control  buttons  will 

cancel the ‘Auto’ function.

When in ‘Auto’ mode, the internal microprocessor 

carries  out  a  battery  voltage  check  each  time  air 

is required and if the voltage becomes too low, the 

‘Auto’ function will be cancelled automatically.

Locking the built-in buttons - fig.  

7

If  desired,  the  built-in  buttons  can  be  locked  to 

prevent them from working.

Hold the battery indicator button (

B

) and then 

the ‘deflate’ button (

C

).

With  buttons  (

B)

  and  (

C)

  held,  press  the 

following buttons in sequence: ‘Auto’ (

D

), ‘Up’ 

(

E

), ‘Auto’ (

D

), ‘Up’ (

E

) - 

All within 3 seconds.

 

Confirmation that the buttons are locked will be 

indicated by a double flash of all the lights.

To  unlock  the  built-in  buttons,  repeat  the  locking 

procedure.

Separate Push-Button Hand Control 

fig.  

8

Connect the hand control (

A

) to the socket (

B

).

Press the ‘Up’  

 button to raise.

 

Press the ‘Down’  

 button to lower.

Summary of Contents for airflo

Page 1: ...user instructions and warranty airflo battery version freedom through lightweight solutions...

Page 2: ...in the presence of flammable anaesthetic mixtures with air oxygen or nitrous oxide PLEASE BE SURE THAT YOU HAVE READ AND UNDERSTOOD THESE INSTRUCTIONS AND THOSE PROVIDED WITH ANY OTHER MANGAR INTERNA...

Page 3: ...re that the battery charger is not connected to the Airflo and then press the battery indicator button A to activate the indicator lights B If the batteries are fully charged all four indicator lights...

Page 4: ...ed If convenient the Airflo may remain continuously on charge when not in use as the batteries cannot be overcharged using this charger WARNING DO NOT recharge the batteries in a bathroom or in a damp...

Page 5: ...AMEL Press the switch to change between Normal and High air flow Normal High WARNING DO NOT operate the Normal High air flow switch while the compressor is running Different ways to operate the Airflo...

Page 6: ...t a battery voltage check each time air is required and if the voltage becomes too low the Auto function will be cancelled automatically Locking the built in buttons fig 7 If desired the built in butt...

Page 7: ...Auto Control Button Move the lever C to up to raise Move the lever to down to lower Move the lever to neutral stop to stop at any height CAMEL and ELK Hand Controls fig 10 Connect the four hoses A fr...

Page 8: ...a new filter a kit of replacement filters is available through your supplier part number CA0233 Battery replacement Turn the Airflo upside down and remove the four screws B followed by the two battery...

Page 9: ...ir flow switch is in the Normal position Auto light flashing This happens when the compressor s internal temperature cut out has tripped Leave the Airflo for 15 minutes to cool The top two lights on t...

Page 10: ...nt AC Fuse with rating Max Min operating environment temperature Auto select button Up button Down button Do Not connect to mains power in a bathroom or outdoors Class II Double Insulated equipment Da...

Page 11: ...or repair by unqualified persons The following statements do not affect the consumer s statutory rights This product is sold on the understanding that in the event of any defect of manufacture or mat...

Reviews: