MP30  |  1.00.01 

11

 

 

N O T E !

 

The  safety  requirements  applicable  to  the  respective  media  to  be 
transported must be observed. 

 

In  order  to  prevent  a  disturbing  accumulation  of  heat,  the  pumps 
should  be  installed  away  from  sources  of  heat  and  freely 
ventilated. 

 

For  installation  outdoors,  the  pump  must  be  installed  in  a 
protective housing frost-free in winter and sufficiently ventilated in 
summer. Direct exposure to sunlight must be avoided.

 

 

 

N O T E !

 

Pumps  have  mechanical  moving  parts  that  can  induce  vibrations. 
To  prevent  damages  at  the  pump  or  at  peripheral  components  / 
facilities  as  well  as  minimizing  noise  development  an  appropriate 
vibration  decoupling  is  necessary.  For  this  M&C  can  deliver  e.g. 
anti-vibration pads.  
This explicit is also valid for the connection of the sample lines at 
the pump head. 

 

 
 
 

11.1 

MECHANICAL 

  The fixing dimensions are shown in Figures 2 and 3. 

  The pumps must be installed in such a way that the impeller is able to draw in sufficient cooling air. 

  The pumps should be installed at the highest point in the system and/or with the pump head 

pointing downward to prevent an accumulation of condensate 

– this will extend the useful life of the 

pump. 

 

 

11.2 

ELECTRICAL 

Electrical installation must take place in compliance with the safety regulations. Before connecting the 
pump, safe isolation from the supply must be ensured. 
 

  The protective conductor must be connected to the pump motor. 

 

  A device for  disconnecting the pump from  the electrical system  must be provided in the electrical 

installation (according to EN 60335-1).

 

 

  The  supply  circuit  should  be  provided  with  a  fuse  with  a  rating  corresponding  to  the  current 

consumption (over-current protection, current consumption, see technical data). 

 
 

  W A R N I N G !

 

 

 

 
The  supply  voltage  must  be  compared  with  the  voltage  shown  on 
the rating plate. A voltage deviation of about 

 10% is admissible.  

 

The  pumps  must  be  mounted  so  that  contact  with  live  parts  (e.g. 
electrical connection, possibly motor windings) is excluded.

 

 

Summary of Contents for MP Series

Page 1: ...Analytical Diaphragm Pump Series MP MP30 Instruction Manual Version 1 00 01...

Page 2: ...sses are to be found in the annexe to this operating manual Please also contact our homepage www mc techgroup com for further information about our products There you can read or download the data she...

Page 3: ...ata 9 9 Dimensions 10 10 Receipt and storage 10 11 Installation information 10 11 1 Mechanical 11 11 2 Electrical 11 11 3 Pneumatic 12 12 Commissioning 13 13 Decommissioning 14 14 Maintenance 14 14 1...

Page 4: ...riate transportation correct storage as well as professional installation and maintenance work All necessary information a skilled staff will need for appropriate use of this product are given in this...

Page 5: ...re opening the equipment it must be switched off and hold no voltages This also applies to any external control circuits that are connected The device is only to be used within the permitted range of...

Page 6: ...minor physical injuries may occur in case the respective safety measures are not fulfilled C AR E Without the warning triangle means that a material damage may occur in case the respective safety mea...

Page 7: ...on proof design The MP30 has a delivery capacity of 7 5l min at atmospheric pressure The output on the pressure side is limited to maximum 2 5 bar absolute Application examples are Transport of gases...

Page 8: ...not be used in areas subject to explosion hazards or for transporting a potentially explosive medium When transporting toxic media system specific and safety relevant regulations must be observed MAC...

Page 9: ...0 Delivery capacity max 7 5 l min Operating pressure 0 14 to max 2 5 bar abs Gas temperature 5 C to 40 C Ambient temperature 5 C to 40 C Storage temperature 15 C to 60 C Gas connections G1 8 i DIN ISO...

Page 10: ...ceipt and check the delivery against the delivery note Check the contents for possible transport damage and immediately notify the transport insurer of any damage NOTE The diaphragm pump should be sto...

Page 11: ...s must be installed in such a way that the impeller is able to draw in sufficient cooling air The pumps should be installed at the highest point in the system and or with the pump head pointing downwa...

Page 12: ...hose couplings the following should be noted NOTE The tightness of the connection can only be ensured when the connecting hose has a straight terminating edge use of a hose cutter Loosen the swivel n...

Page 13: ...king pressure of the pump is not exceeded With the pump stationary normal atmospheric pressure must be established in the pipes Diaphragms and valve plates are the only wearing parts of the pumps Wear...

Page 14: ...process specific safety measures must be observed W AR N I N G Dangerous voltage Before working on the diaphragm pump the same must be safely isolated from the supply and disconnection from the suppl...

Page 15: ...making a continuous line with a felt tip pen Loosen four head screws 5 Remove pressure plate 15 head cover 4 and intermediate plate 1 from pump housing Loosen four fixing screws in impeller cover 13 a...

Page 16: ...on by lightly moving the valve seat Fit sealing rings 3 in intermediate plate 1 Join head cover 4 and intermediate plate 1 together felt tip pen mark must be in alignment and check centred seat throug...

Page 17: ...rheating Delivery capacity pressure or vacuum too low Diaphragms or valve plates worn Replace worn parts see 14 ff Problem Indication Possible cause Check Remedy Initially compare reached pump capacit...

Page 18: ...nd on specific operating conditions The recommended quantities are based on experience and are not binding Diaphragm Pump Type MP30 C consumable parts S spare parts Recommended quantity MP30 being in...

Reviews: