Making measurements - a tutorial
Chapter 4
Mastersizer 2000
Page 4-25
Measurement Cycles dialogue
Use the
Measurement
Cycles
dialogue to set up repeat measurements and set up
tank cleaning and dispersant degassing. This diagram shows the features:
ill 7760
Cleaning
It is possible to set up the SOP so that the tank and cell are cleaned together:
Before each aliquot.
After each aliquot.
Before
and
after each aliquot (select both check boxes).
Selecting the relevant
Enable
box causes a clean sequence to be performed.
If the sample/dispersant used is not normally removed with a single flush, enter a
number in the
Flush cycles
box to run the flush sequence that number of times.
The user can set the
Clean mode
to either
Automatic
or
Manual
.
Manual
ena-
bles the user to clean the system easily using a beaker and a bottle of clean solvent.
A series of prompts guides the user through the cleaning cycle.
Dispersant degassing
Selecting the
Enable
check box causes a degassing routine to be performed each
time the tank is cleaned. The degassing routine uses ultrasonic action to help
remove dissolved gases from the dispersant.
1
2
MAN0384-1.0 Mastersizer 2000.book Page 25 Tuesday, March 20, 2007 8:56 AM
Summary of Contents for Mastersizer 2000
Page 2: ......
Page 6: ...Page 2 Addendum 0070 2 0 Addendum 0070 2 0 fm Page 2 Wednesday March 10 2010 11 58 AM ...
Page 14: ...MAN0384 1 0 Mastersizer 2000 book Page 2 Tuesday March 20 2007 8 56 AM ...
Page 124: ...MAN0384 1 0 Mastersizer 2000 book Page 2 Tuesday March 20 2007 8 56 AM ...
Page 133: ...Part 3 Appendices MAN0384 1 0 Mastersizer 2000 book Page 1 Tuesday March 20 2007 8 56 AM ...
Page 134: ...MAN0384 1 0 Mastersizer 2000 book Page 2 Tuesday March 20 2007 8 56 AM ...
Page 153: ......
Page 154: ......