Chapter 5
Maintenance
Page 5-2
MAN 0494
5
Refer to the following illustration for confirmation.
ill 8587
1.
Use an allen key to remove all four bolts
and
from the transit locks, pic-
tured above, then remove the angle brackets.
2.
Move the needle holder approximately 80mm left, to the centre of the tray, by
applying light finger pressure.
3.
Switch on the instrument and then start the software. The needle will move to
the wash station.
4.
Restart the software to initialise the needle to its correct position.
5.
Retain all bolts and brackets for future shipment of the instrument.
Contact the Malven Helpdesk for guidance on the procedure of refitting the
transit locks.