www.malmsten.com
Malmsten AB • www.malmsten.com
+46 44 28 71 00 • [email protected]
v.47
-23E
MAINTENANCE
11
Swimming Pool Lift
DEVICE
i-swim2
MODEL
Page
32
of
38
11.2 PERIODICAL INSPECTION
PERIODICAL INSPECTION MUST BE CARRIED OUT ONLY BY
SKILLED PERSONNEL.
Every six months, a periodical inspection of the lift has to be carried out by a technician properly trained.
The inspections consist of a
visual examination
for the control of:
1. Supporting structure;
2. Lifting device with relevant connections;
3. Brakes;
4. Controls, emergency stop button;
5. Safety devices;
6. Support devices for the person (seat belt).
The inspections also include:
7. Operation tests;
8. Maintenance;
9. Adjustments.
10. Each periodical inspection shall include a test with the workload of one (1) lifting cycle with maximum
load (simulated).
Each remark for the safety of the lift shall be entered on a register (see facsimile in Annex) that the user must
keep in a safe place for the examination in the event of accident.
The record of the inspection shall also contain the following information:
- Date of the inspection;
- Identification details and serial number of the machine;
- Date of next inspection;
- Identification and signature of the person in charge for the inspection.
If the inspection points out any defect, signs of wear or other damages that compromise the safety of the lift. In
the event of immediate danger for the safety, the user must put out of service the lift until the fault has been
solved.
DEFECTS OR DAMAGE MUST BE REPORTED TO THE
MANUFACTURER AUTOLIFT SRL FOR APPROPRIATE ACTION.