Bedpan / Urinal Bottle and Utensil / Bowl Washer Disinfector (WDS)
Operation, Maintenance and Installation Manual
Issue 16
Page
12
August 22
2.3 Plumbing
These installation guidelines must be followed to ensure the device will operate as intended.
Installations must be carried out by a qualified, licensed tradesperson.
Service Connections
NOTE: Plumbing connection must comply with AS/NZS 3500.1 and be Watermark Certified.
Waste Connection
•
HOT AND COLD WATER CONNECTIONS ARE REQUIRED. The device can be connected to any potable mains pressure
hot and cold water supply as a back flow prevention air gap is incorporated in the design of the water tank. Complies to
AS 2845.2
•
The water supply is to be connected to an isolating valve or cistern stopcock placed approximately 1200mm from the
floor to the right-hand side of the device (preferably not behind the device).
Cold water: Flow pressure 100-350Kpa max. Temperature 15° - 25°C
Hot water: Flow pressure 100-350Kpa max. Temperature 55° - 60°C
Minimum water flow for 100Kpa 9.377L/min
If inlet water flow pressure is higher than 350 Kpa an inline pressure reducing valve should be fitted.
Soil Line
•
Hot water. Discharge temp 75°
–
80°C. Soil line must comply to AS/NZS 3500.2
•
Soil line connection is by a pan collar or other preferred method. If the belled end on the polyethylene moulded trap is
not required it can easily be cut off to provide a straight pipe connection.
•
Level the device by using the flanged screw in legs and, if possible, maintain approx. 100mm floor clearance for ease of
floor cleaning. Malmet recommends affixing some of the leg flanges to the floor via stainless self-tapping screws to
prevent sideways movements and damage to services and soil line connections.
•
The soil line should protrude from the floor or wall at a minimum of 100mm.
2.4 Electrical
These installation guidelines must be followed to ensure the device will operate as intended and must be carried out by
a qualified, licensed electrical tradesperson. The device must be installed and serviced to national wiring rules AS/NZS
3000
.
Model:
WDS 1ph
240V 50Hz
1 ph
20Amp
WDS 3ph
415V 50Hz
3 ph
20Amp
Overcurrent Protection device: A 20A circuit breaker or fuse must be installed in the building installation for
the device.
•
Devices are supplied with 1700mm power supply cord extending from the rear top right hand side of the device for hard
wiring into Mains ISO switch. Position switch approx. 1500mm above floor level.
•
An ISO switch or circuit breaker must be included in the installation (not supplied by Malmet).
•
Mains power ISO switch must be suitably located and easily reached, approximately 1500mm above floor level adjacent
to device. The 1.7m mains power lead exits the device approximately 1600mm above floor level on the right hand side of
the device.
•
Mains ISO switch must be marked as the disconnecting device for the equipment.
•
Mains ISO switch must have contact disconnection of all poles to provide full disconnection.