background image

Fluid Warming Cabinet

 

Operation, Maintenance and Installation Manual

 

Issue 18 

Page 7

 

Jun-23

 

2.0 Installation 

Floor Bench Skirt 

Floor Bench Skirt is factory fitted to the cabinet.  To level, remove the four plastic plugs from the inside floor and 
adjust the feet by turning the slotted screw head. Once the device is level, replace the plugs. 

If  the  Floor-Bench  Skirt  is  ordered  separately  to  convert  a  warming  cabinet,  the  conversion  will  take 
approximately fifteen minutes and requires a screwdriver only.  Carefully, place the warming cabinet on its back.  
The base is pre-drilled with ten Nutsert holes.  Using the screws supplied, fasten the Floor-Bench Skirt.  Carefully, 
stand the warming cabinet upright and level as above. 

The front panel of the Floor-Bench Skirt is removable to clean under the warming cabinet. 

Combination Cabinets 

To  level,  remove  the  four  plastic  plugs  from  the  inside  floor  and  adjust  the  feet  by  turning  the  slotted  screw 
head. Once the warming cabinet is level, replace the plugs. 

The front panel of the floor skirt is removable to clean under the warming cabinet. 

Cabinet Stand 

Cabinet Stand is factory fitted to the warming cabinet.  To level, remove the front panel of the stand and adjust 
the feet by turning the slotted screw-head.  This panel can also be removed to clean under the warming cabinet. 

If the stand is ordered separately to convert a warming cabinet, the conversion will take approximately fifteen 
minutes  and  requires  a  screwdriver  only.    Carefully,  place  the  warming  cabinet  on  its  back.    The  base  is  pre-
drilled with ten Nutsert holes.  Using the screws supplied, fasten the stand.  Carefully, place the warming cabinet 
upright and level as above. 

Combination Conversion Kit 

This  comprises  a  Floor-Bench  Skirt  and  a  Rear  Joining  Bracket.    Conversion  will  take  approximately  fifteen 
minutes and requires a screwdriver only.  Carefully, place the bottom warming cabinet on its back.  The base is 
pre-drilled with ten Nutsert holes.  Using the screws supplied, fasten the Floor-Bench Skirt.  Carefully, stand the 
warming cabinet upright. 

Remove  the  7  plugs  (3  small  and  4  large)  on  the  top  of  the  bottom  warming  cabinet,  place  the  top  warming 
cabinet  in  position  locating  the  bottom  hinge  bolts  in  the  four  large  holes.    Remove  the  two  screws  in  the 
warming cabinet that secure the slide out service tray and slide out as far as it will go.  There are three securing 
holes located inside the top of the service compartment.  Using the screws supplied, fasten the warming cabinets 
together and replace the service tray. 

The back of both warming cabinets is pre-drilled with four Nutsert holes, using the screws supplied, fix the Rear 
Joining Bracket into position.  Place the warming cabinet in position and level. 

 

Plugs must be replaced after levelling or the temperature control of the warming cabinet may be compromised

Wall Mounted Cabinet 

Wall  Mounting  Brackets  are  supplied  for  single  210  litre  and  260  litre  models. Secure  the  bracket  to the  wall, 
ensuring it is level.  It is recommended to use 4 (8mm x 50mm)  long coach screws for stud position walls and 
dyna bolts for solid walls. 

Two people are required to lift the cabinet to the height of the Mounting Bracket, move the top of the warming 
cabinet  to  the  wall  at  approximately  45°  and  locate  on  the  Wall  Mounting  Bracket.    When  secure  swing  the 
bottom of the warming cabinet to the wall ensuring it is firmly pushed onto the wall bracket.  To remove, reverse 
this procedure. 

CAUTION:  Wall mounted fluid cabinet when fully loaded weighs approximately 

143 kg

. unloaded weight approximately 

83 

kg

  

Summary of Contents for F105BT

Page 1: ...state Issue 18 Jun 23 Head Office and Customer Service ABN 95 001 717 791 9 11 McKay Avenue PO Box 373 Leeton NSW 2705 Phone 61 2 6953 7677 Email info malmet com au Fluid Warming Cabinet Single Model...

Page 2: ...rs 4 1 1 Device Operation 5 1 2 Control Display Features 5 1 3 Operating Features 5 1 4 Changing the set temperature lock out feature 6 1 5 Temperature Cut Out 6 1 6 Temperature Sensor Plate 6 2 0 Ins...

Page 3: ...wever owing to Malmet Australia s policy of continuous improvement changes to these specifications may be made at any time without notice on the part of Malmet Australia Pty Ltd Certifications and Com...

Page 4: ...be replaced by a special cord or assembly from the manufacturer or its service agent Do not overload shelves Do not obstruct hot air outlet above top shelf Castors where fitted are for convenience bu...

Page 5: ...able 15 litres per shelf maximum load Single 420 Litre Fluid Warming Cabinets Models F420FS and F420C are designed to hold a maximum of 105 litres of fluid in 1 and 2 litre bottles and bags 420 litre...

Page 6: ...age circuit breaker for approximately three hours This will allow any moisture in the heater to dry out The device can then be connected to an earth leakage circuit breaker protected circuit if requir...

Page 7: ...c cut out fails a secondary safety bi metallic cut out temperature switch will switch off the power to the elements Fig 1 1 6 Temperature Sensor Plate When new product is introduced into the cabinet t...

Page 8: ...Combination Conversion Kit This comprises a Floor Bench Skirt and a Rear Joining Bracket Conversion will take approximately fifteen minutes and requires a screwdriver only Carefully place the bottom w...

Page 9: ...s wikimedia org w index php curid 55610678 Handling Weights of Device Net 47 195 kg Shipping 57 210 kg Shipping with crating 84 275 kg Handling of the device to installation site must be with fork lif...

Page 10: ...ration Maintenance and Installation Manual Issue 18 Page 9 Jun 23 2 1 Service Connections Electrical Connection 240 Volt 10 Amp Plug into 10 Amp G P O 2 2 Device Dimension and Clearances F105BT Fluid...

Page 11: ...Fluid Warming Cabinet Operation Maintenance and Installation Manual Issue 18 Page 10 Jun 23 F210FS Fluid Warming Cabinet Freestanding with Floor Skirt...

Page 12: ...Fluid Warming Cabinet Operation Maintenance and Installation Manual Issue 18 Page 11 Jun 23 F260FS Fluid Warming Cabinet Freestanding with Floor Skirt...

Page 13: ...not intended for use by persons including children with reduced physical sensory or mental capabilities or lack of experience and knowledge Children should be supervised to ensure they do not play wit...

Page 14: ...her authorised personnel 1 Check electrical connections as per AS NZS 3551 2012 2 Perform temperature calibration Insert calibrated thermometer Allow cabinet temperature to stabilise Compare cabinet d...

Page 15: ...r Service Display turns on Runs for approximately 1 hour then turns off Fan failure Switch off power supply at main and Call for Service Display is on but unit not heating Element failure Relay board...

Page 16: ...Phase Hz 1 ph 50Hz Amps 2 8 Amps Watts 0 6 kW Environment operating conditions Temperature Relative Humidity 10 C to 25 C 30 to 70 Electrical Connection IEC right angle Power Cord with 3 Pin Plug 10 A...

Page 17: ...Fluid Warming Cabinet Operation Maintenance and Installation Manual Issue 18 Page 16 Jun 23 4 2 Wiring Diagram...

Page 18: ...for issuing the relevant certificates of compliance these may differ from state to state CONDITIONS AND EXCLUSIONS Device must be installed and commissioned according to Malmet s instructions outline...

Page 19: ...ther way from the purchaser to any other person To the extent permitted by law unauthorised use of any parts that were not supplied or approved for use in the applicable device by Malmet will result i...

Page 20: ...component under this warranty Please complete details below Date Purchased Warranty Expiry Date Sold To For Service Contact PROOF OF PURCHASE Please retain your proof of purchase receipt invoice or c...

Page 21: ...Fluid Warming Cabinet Operation Maintenance and Installation Manual DCN 2970 P N 8 M00 84 900 Manufactured by Malmet Australia Pty Ltd ABN 95 001 717 791 www malmet com au...

Reviews: