1.4 Changing the set temperature (lock out feature)
AUTHORISED PERSONNEL ONLY
Authorised operators can change the temperature by holding the up and down temperature buttons at the same
time for a total of 5 seconds. The temperature can then be adjusted. The setting will revert back to tamper
proof 5 seconds after the temperature is adjusted or if the up and down temperature buttons are not pressed
within 5 seconds.
1.5 Temperature Cut Out
The device has a built in electronic over temperature cut out which will switch off the elements if the selected
temperature is exceeded by 5°C, and ‘Ot’ will flash o
n the display panel. (See Fig 1)
If the electronic cut out fails, a secondary safety bi-metallic cut out temperature switch will switch off the power
to the elements.
Fig 1
1.6 Temperature Sensor Plate
When new product is introduced into the cabinet the temperature of the fluid can be monitored using the
temperature sensor plate.
1.
Place fluid container on the Fluid Sensor Plate located on the bottom shelf ensuring the sensor disc is
completely covered.
2.
Wait two minutes for an accurate temperature reading to be obtained.
3.
To view actual fluid temperature, via the main display board, press and hold down either of the up or
down temperature buttons.
Note 1: It can take an average of eight hours for a full load of fluid containers to reach required temperature.
This time may vary depending on the configuration of the load. It is recommended that the cabinet is
re-stocked at the end of each day.
Note 2: Rotation of stock is important, as once warmed, solution should be used within 14 days.