Positioning of Bedpan Washer Disinfector
Freestanding Model
Model
Placement
Access
Required
Unit Dimensions
Height
Width
Depth (mm)
ES910
ES930
Freestanding
Both Sides
1305
595
660
For Freestanding model, please allow sufficient room for servicing purposes. Recommended space
requirements 200mm
on either side and 150mm at the rear of the unit.
New Buildings
Service connections are normally pre-placed after planning and consultation with all interested parties.
Installation is by connection to the services provided.
As the soil line (sewerage outlet) is the least flexible of all the connections, this usually influences the
decision as to where to place the Bedpan Washer Disinfector.
The Bedpan Washer Disinfector
is supplied with either a ‘S’ or ‘P’ Trap as n
ominated at the time of
purchase
. The ‘S’ Trap connects through the floor and the ‘P’ Trap connects through the back wall. The
trap section is easily removed if the wrong trap has been ordered. Refer to Diagram C2 for trap
connections.
Potential electromagnetic or other interference between other EQUIPMENT and other devises can
possibly affect the Infra red hands free operation sensor. It is advisable to check all the equipment
and devices in the intended installation area that have infra red operation.
Electromagnetic
interference can be prevented by installing the device in non patient areas of hospital (or
similar).