Bedpan Washer Disinfector - Energy Saver
Detergent Model With Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 21
Page 22
14 September 2017
Fault 07
Hot Water Tank Low Water Level Sensed During Cycle
Cause
Faulty water level probe
Cold water tank empty or not filling quick enough
Leaking tank or hose
Element stuck on
Condensing coil blocked
Fault 08
Door Closure Fault
–
Not Confirmed Door Open
(door switch still indicates closed)
Cause
Door limit switch stuck on
Fault 09
Failed To Reach Disinfection Start Temperature Resulting In Time Out
Cause
Steam leak
Final rinse solenoid stuck open
Faulty temperature sensor hot water tank
Blown element leg
Steam condensate leaking cold water into chamber
Faulty temperature sensor chamber
Door not reaching interlocking micro switch. Door not fully closed.
Over Temperature Board (Power to element open circuit)
Element overtemp protection manual reset thermal cut-out switch open
circuit.
Fault 10
Disinfection Temperature Not Maintained Resulting In Cycle Time Out
Cause
Steam leak
Final rinse solenoid stuck open
Faulty temperature sensor hot water tank
Blown element leg
Steam condensate leaking cold water into chamber
Fault temperature sensor
–
chamber
Fault 12
Hot Tank Short Temperature Sensor
Fault 13
Hot Tank Open Circuit Temperature Sensor
Fault 14
Disinfection Chamber Short Circuit Temperature Sensor
Fault 15
Disinfection Chamber Open Circuit Temperature Sensor
Cold Water Overflow Indication
The Yellow LED is used for Cold Tank Overflow
.
Overflow is indicated by a STEADY ON LED, whenever the overflow level is detected.