56
Part No. 20751 Rev-01
2014-08
High Calibration
Connection Diagrams
Connection Diagram for NO and NO
2
High
Range Calibration
1. Cylinder Cap
2. Regulator Seal
3. Regulator
4. Tubing Adapter
5. Attach Tube Kit
6. Turn valve counter-clockwise to
VWDUWJDVÀRZ
7
To
INOmax DS
IR
6a. If the pressure is in the red or black
zone (0-25 psig) select another INOcal
cylinder.
7. Attach tube kit to sample line
8. Calibrate sensor
9. Remove sample line from tube kit
7XUQYDOYHFORFNZLVHWRVWRSJDVÀRZ
11. One-way Valve
6
6a