WARNING:
• A new INOMAX cylinder and regulator must be purged
before use to ensure the patient does not receive an excess
level of NO
2
.
/RVVRIFRPPXQLFDWLRQEHWZHHQWKH,12PD['6
IR
and the
INOMAX cylinder for more than one hour will result in
interruption of INOMAX delivery.
Caution: When using the Transport Regulator/Cap Assembly
(PN 10022) ensure the cap is in place on the cylinder
and the infrared cable is connected to the infrared
connector port on the back of the INOmax DS
IR
.
Note: Check the INOMAX cylinder for the correct product identity
labels, cylinder concentration and expiration date. Ensure
the INOMAX gas cylinder has more than 500 psig.
Step One
Note: Ensure the white plastic tip is in place on the regulator
connector and not chipped or cracked (see Figure 2).
Connect a high
pressure regulator
to an INOMAX
cylinder and tighten
WKH¿WWLQJWRWKH
INOMAX cylinder
(see Figure 1).
Summary of Contents for INOmax DSIR
Page 6: ...6 Part No 20751 Rev 01 2014 08 Pre Use Checkout...
Page 7: ...7 Part No 20751 Rev 01 2014 08 Pre Use Checkout Intentionally left blank...
Page 21: ...21 Part No 20751 Rev 01 2014 08 Circuit Connection Diagrams Intentionally left blank...
Page 45: ...45 Part No 20751 Rev 01 2014 08 Circuit Connection Diagrams Intentionally left blank...
Page 55: ...55 Part No 20751 Rev 01 2014 08 Changing INOMAX Cylinders Intentionally left blank...
Page 60: ...60 Part No 20751 Rev 01 2014 08 0DOOLQFNURGW 0DQXIDFWXULQJ HPULWH ULYH 0DGLVRQ 86...