Part No. 20732 Rev-02
2014-02
15
Cautions:
• When cleaning, do not submerge in liquid.
• Do not autoclave.
Cleaning Procedure
Maintenance
Cleaning Agent
Active Ingredients
Precise Hospital Foam Cleaner
Disinfectant
by Caltech Industries
o-Phenylphenol < 0.37%
Other ingredients 99.63%
Pure Green 24
by Pure Green, LLC
SDC – silver ions 0.003%
Citric acid 4.84%
Other ingredients 95.157%
PDI Super Sani Cloth
by PDI
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chlorides 0.25%
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium
chlorides 0.25%
Isopropyl alcohol 55%
Inert ingredients 44.50%
Sani Cloth HB
by PDI
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium
chlorides 0.07%
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chlorides 0.07%
Inert ingredients 99.86%
Asepti-HB
by Ecolab Inc.
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium
chlorides 0.07%
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chlorides 0.07%
Inert ingredients 99.86%
Cavicide and CaviWipes
by Metrex
Diisobutylphenooxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%
Isopropyl alcohol 17.2%
Inert ingredients 82.52%
Caution
:
• Apply cleaning agent to a cloth before application; do not spray cleaning agent directly on the device.
It is important to prevent pooling and direct contact with open connections which can cause damage
over time.
- Clean the outer surface of the INOblender with a soft cloth dampened in a mild soap and water solution
(Example: Palmolive Ultra antibacterial, 4 drops per 50 mL of warm water), isopropyl alcohol (70%), or
with one of the following cleaning agents while following the manufacturer’s recommendations.
Note: The INOblender does not contain any user repairable parts.
Bioquell Hydrogen Peroxide Sterilant
Bioquell Hydrogen Peroxide Sterilant and hydrogen peroxide vapor generators are regulated by the US
Environmental Protection Agency (EPA) as pesticide chemicals in accordance with Federal Insecticide,
Fungicide, and Rodenticide Act (FIFRA). Use of these products in cleaning/disinfection processes have not
been validated with the INOblender. Do not use these products to decontaminate the INOblender or any
ancillary products used with the INOblender.
Summary of Contents for INOblender 10026
Page 1: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 Operation and Maintenance Manual English...
Page 4: ...Part No 20732 Rev 02 2014 02 2 Intentionally left blank...
Page 5: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 1 General Information General Information...
Page 6: ...Part No 20732 Rev 02 2014 02 1 General Information General Information...
Page 11: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 2 Setup Setup...
Page 12: ...Part No 20732 Rev 02 2014 02 2 Setup Setup...
Page 14: ...Part No 20732 Rev 02 2014 02 8 Intentionally left blank...
Page 15: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...
Page 16: ...Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...
Page 18: ...Part No 20732 Rev 02 2014 02 10 Intentionally left blank...
Page 19: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 4 Operation Operation...
Page 20: ...Part No 20732 Rev 02 2014 02 4 Operation Operation...
Page 21: ...Part No 20732 Rev 02 2014 02 11 Intentionally left blank...
Page 24: ...Part No 20732 Rev 02 2014 02 14 Intentionally left blank...
Page 25: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...
Page 26: ...Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...
Page 32: ...Part No 20732 Rev 02 2014 02 20 Intentionally left blank...
Page 33: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 6 Product Specifications...
Page 34: ...Part No 20732 Rev 02 2014 02 6 Product Specifications...
Page 36: ...Part No 20732 Rev 02 2014 02 22 Intentionally left blank...
Page 37: ...17 Part No 20732 Rev 02 2014 02 23 Intentionally left blank...