Part No. 20732 Rev-02
2014-02
18
Cylinder Leak Check
If a leak is suspected during the high-pressure leak test (see Section 3/Pre-Use Checkout; High Pressure
Leak Test), the following steps can be taken to check for leaks (see Figure 5-3 for possible cylinder gas leak
locations) in the INOMAX regulator or INOMAX cylinder.
Figure 5-3
1. Cylinder Valve Regulator Connection
2. INOMAX Regulator Hand Wheel Connection
3. Regulator End Cap Connection
4. Tamper Evident Tape
5. Valve Nut
6. Safety Pressure Release Device
1. Ensure that INOMAX regulator is connected to
cylinder valve outlet (hand-tighten only), that the
cylinder valve is open, and that the cylinder has
more than 14 bar (200 psig).
2. Apply soapy water to points 1, 2, 3, 5 and 6 (see
Figure 5-3); if bubbles form, there is a leak.
3. If there are no bubbles, leak may be inside
INOblender or INOmax DS
IR
and cannot be
repaired. Replace INOblender or INOmax DS
IR
and contact Customer Support.
Recommended actions should a leak be detected:
1. A leak detected at points 1 and 2 (see Figure
5-3) may be corrected by tightening the INOMAX
regulator hand wheel.
a. If cylinder valve is open, close cylinder valve
and tighten INOMAX regulator hand wheel.
b. Open cylinder valve and reapply soapy water
to points 1 and 2.
c. If bubbles form, there is a leak.
d. Remove INOMAX regulator and check white
plastic tip on INOMAX regulator for chips or
cracks. Replace if necessary (see replacing
the tip on the INOMAX regulator, page 16).
For the ISO type regulator connector, check
that the O-ring is present and is not damaged.
Replace if necessary (see replacing the
O-ring on the INOMAX regulator, page 17).
Repeat step b (note: If leak remains, replace
INOMAX regulator).
2. If a leak is detected between the regulator body
and regulator end cap (see point 3), replace
INOMAX regulator and contact Customer Support.
3. A leak detected at cylinder valve nut connection
(see point 5) may not be repaired.
Replace INOMAX cylinder and contact Customer
Support.
4. A leak detected at safety pressure release device
(see point 6) may not be repaired. Replace
INOMAX cylinder and contact Customer Support.
6
5
1
2
3
4
Note: Refer to hospital policies and procedures
for dealing with leaking gas cylinders.
Additional information regarding
environmental effects can be found in
the section 1/General Information.
Summary of Contents for INOblender 10026
Page 1: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 Operation and Maintenance Manual English...
Page 4: ...Part No 20732 Rev 02 2014 02 2 Intentionally left blank...
Page 5: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 1 General Information General Information...
Page 6: ...Part No 20732 Rev 02 2014 02 1 General Information General Information...
Page 11: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 2 Setup Setup...
Page 12: ...Part No 20732 Rev 02 2014 02 2 Setup Setup...
Page 14: ...Part No 20732 Rev 02 2014 02 8 Intentionally left blank...
Page 15: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...
Page 16: ...Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...
Page 18: ...Part No 20732 Rev 02 2014 02 10 Intentionally left blank...
Page 19: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 4 Operation Operation...
Page 20: ...Part No 20732 Rev 02 2014 02 4 Operation Operation...
Page 21: ...Part No 20732 Rev 02 2014 02 11 Intentionally left blank...
Page 24: ...Part No 20732 Rev 02 2014 02 14 Intentionally left blank...
Page 25: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...
Page 26: ...Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...
Page 32: ...Part No 20732 Rev 02 2014 02 20 Intentionally left blank...
Page 33: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 6 Product Specifications...
Page 34: ...Part No 20732 Rev 02 2014 02 6 Product Specifications...
Page 36: ...Part No 20732 Rev 02 2014 02 22 Intentionally left blank...
Page 37: ...17 Part No 20732 Rev 02 2014 02 23 Intentionally left blank...