Cylinder Leak Check

If a leak is suspected during the high-pressure leak test (see Section 3/Pre-Use Checkout; High Pressure 

Leak Test), the following steps can be taken to check for leaks (see Figure 5-3 for possible cylinder gas leak 

locations) in the INOMAX regulator or INOMAX cylinder.

Figure 5-3

1.  Cylinder Valve Regulator Connection

2.  INOMAX Regulator Hand Wheel Connection

3.  Regulator End Cap Connection

4.  Tamper Evident Tape

5.  Valve Nut

6.  Safety Pressure Release Device

1.  Ensure that INOMAX regulator is connected to 

cylinder valve outlet (hand-tighten only), that the 

cylinder valve is open, and that the cylinder has 

more than 14 bar (200 psig).

2.  Apply soapy water to points 1, 2, 3, 5 and 6 (see 

Figure 5-3); if bubbles form, there is a leak.

3.  If there are no bubbles, leak may be inside 

INOblender or INOmax DS

IR

 and cannot be 

repaired. Replace INOblender or INOmax DS

IR

 

and contact Customer Support.

 

Recommended actions should a leak be detected:

1.  A leak detected at points 1 and 2 (see Figure 

5-3) may be corrected by tightening the INOMAX 

regulator hand wheel.

a. If cylinder valve is open, close cylinder valve 

and tighten INOMAX regulator hand wheel.

b. Open cylinder valve and reapply soapy water 

to points 1 and 2.

c. If bubbles form, there is a leak.
d. Remove INOMAX regulator and check white 

plastic tip on INOMAX regulator for chips or 

cracks. Replace if necessary (see replacing 

the tip on the INOMAX regulator, page 16). 

For the ISO type regulator connector, check 

that the O-ring is present and is not damaged. 

Replace if necessary (see replacing the 

O-ring on the INOMAX regulator, page 17). 

Repeat step b (note: If leak remains, replace 

INOMAX regulator).

2.  If a leak is detected between the regulator body 

and regulator end cap (see point 3), replace 

INOMAX regulator and contact Customer Support.

3.  A leak detected at cylinder valve nut connection 

(see point 5) may not be repaired.

 

Replace INOMAX cylinder and contact Customer  

Support.

4.  A leak detected at safety pressure release device 

(see point 6) may not be repaired. Replace 

INOMAX cylinder and contact Customer Support.

6

5

1

2

3

4

Note:   Refer to hospital policies and procedures  

for dealing with leaking gas cylinders.

 

Additional information regarding

 

environmental effects can be found in 

 

the section 1/General Information.

Summary of Contents for INOblender 10026

Page 1: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 Operation and Maintenance Manual English...

Page 2: ...e faulty maintenance improper repair damage or alteration by anyone other than Mallinckrodt Representatives Outside the U S A and Canada check local laws for any restrictions that may apply Caution U...

Page 3: ...roduction 4 Definitions and Abbreviations 4 Theory of Operation 5 Environmental Effects 6 2 Setup 7 3 Pre Use Checkout 9 4 Operation 12 5 Maintenance 15 Replacing Regulator Tips 16 Cylinder Leak Check...

Page 4: ...Part No 20732 Rev 02 2014 02 2 Intentionally left blank...

Page 5: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 1 General Information General Information...

Page 6: ...Part No 20732 Rev 02 2014 02 1 General Information General Information...

Page 7: ...y sensitive to NO or NO2 or who may be exposed to these agents for prolonged periods as a result of the use of this device should be aware that the INOblender does not scavenge the exhaust drug and th...

Page 8: ...to be mixed into a user set flow of oxygen which is delivered to a patient The intended use for the INOblender is as a back up to a primary INOMAX delivery system or for short term attended use when a...

Page 9: ...INOblender is also connected to a 2 bar 30 psig INOMAX 800 ppm or 400 ppm supply The INOMAX flow is controlled by a balance regulator in proportion to the oxygen pressure caused by the fixed restrict...

Page 10: ...easurements as performed by Hess et al The NO and NO2 concentrations were measured using a chemiluminescence analyzer when 100 ppm of NO at 8 L min was delivered into a room with no scavenging being u...

Page 11: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 2 Setup Setup...

Page 12: ...Part No 20732 Rev 02 2014 02 2 Setup Setup...

Page 13: ...pply to the O2 inlet fitting see part list for correct oxygen hose Note 3 5 bar 50 psig nominal 1 Make sure the NO dose setting dial is turned to zero 2 Attach a manual resuscitation bag to the INOble...

Page 14: ...Part No 20732 Rev 02 2014 02 8 Intentionally left blank...

Page 15: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Page 16: ...Part No 20732 Rev 02 2014 02 3 Pre Use Checkout Pre Use Checkout...

Page 17: ...nnot be traced replace the INOblender Delivery Confirmation and Purge 1 Set the INOblender to 40 ppm when using an 800 ppm cylinder 20 ppm when using a 400 ppm cylinder 2 Set the oxygen flow on the IN...

Page 18: ...Part No 20732 Rev 02 2014 02 10 Intentionally left blank...

Page 19: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 4 Operation Operation...

Page 20: ...Part No 20732 Rev 02 2014 02 4 Operation Operation...

Page 21: ...Part No 20732 Rev 02 2014 02 11 Intentionally left blank...

Page 22: ...must be followed to help ensure NO2 is purged from the system before the manual resuscitator bag is connected to the patient The manual bag should be squeezed repeatedly during use to avoid NO2 from...

Page 23: ...1 Turn the INOblender setting dial to the desired concentration 5 to 80 ppm for an 800 ppm cylinder and 2 5 to 40 ppm for a 400 ppm cylinder 2 Turn the O2 flowmeter to the desired flow rate 5 to 14 L...

Page 24: ...Part No 20732 Rev 02 2014 02 14 Intentionally left blank...

Page 25: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Page 26: ...Part No 20732 Rev 02 2014 02 5 Maintenance Maintenance...

Page 27: ...by Metrex Diisobutylphenooxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropyl alcohol 17 2 Inert ingredients 82 52 Caution Apply cleaning agent to a cloth before application do not spray c...

Page 28: ...SO see Figure 5 2 by pulling on the tip a and turning it counterclockwise b see Figure 5 1 3 Ensure the threads are clean on the regulator tip if required use a lint free cloth 4 Install the new tip F...

Page 29: ...the O ring as they may damage the metal gland and cause a leak O ring Gland Replacing the O ring on the ISO 5145 INOMAX regulator fitting Note Depressurize the INOMAX regulator by using the purge port...

Page 30: ...tightening the INOMAX regulator hand wheel a If cylinder valve is open close cylinder valve and tighten INOMAX regulator hand wheel b Open cylinder valve and reapply soapy water to points 1 and 2 c If...

Page 31: ...1 INOMAX Regulator CGA 626 10006 INOMAX Regulator ISO 5145 10020 INOMAX Regulator Extension Hose 10014 Operation and Maintenance Manual 20732 O ring ISO 5145 INOMAX Regulator 80470 Tip CGA 626 INOMAX...

Page 32: ...Part No 20732 Rev 02 2014 02 20 Intentionally left blank...

Page 33: ...Part No 20003 Rev 01 Part No 20732 Rev 02 2014 02 6 Product Specifications...

Page 34: ...Part No 20732 Rev 02 2014 02 6 Product Specifications...

Page 35: ...ting and Storage Conditions Operating Temperature 10 to 40 C Operating Humidity 15 to 95 non condensing Operating Barometric Pressure 57 to 110 kPa Storage Temperature 40 to 70 C Storage Humidity 15 t...

Page 36: ...Part No 20732 Rev 02 2014 02 22 Intentionally left blank...

Page 37: ...17 Part No 20732 Rev 02 2014 02 23 Intentionally left blank...

Page 38: ...565 Europe European Authorized Representative MDSS GmbH Schiffgraben 41 30175 Hannover Germany Part No 20732 Rev 02 2014 02 Manufacturer North America Mallinckrodt Manufacturing LLC 6603 Femrite Driv...

Reviews: