background image

The motor and phase rotation meter uses a 9V battery. To replace the 
battery, follow these steps.
1. 

Place the motor and phase rotation meter face down on a non-abrasive 
surface and loosen the battery-door screw with a screwdriver.

2. 

Lift the battery access lid away from the meter

3. 

Observe the battery polarity shown in the battery compartment.

4. 

Secure the battery access lid back in position with the screw.

Safety Information

 

Caution identifies conditions and actions that may damage the 

MT935. Warning identifies conditions and actions that pose 
hazard to the user.

Read First: Safety Information

To avoid possible electric shock or fire, do the following:

Read the following safety information carefully  before using or servicing 

the instrument.

Adhere to local and national safety codes.

Protective equipment must be used  to prevent shock and injury. 

Use of instrument in a manner not specified by  the manufacturer may 

impair safety features/protection provided by the equipment.

Avoid working alone.

When using the probes, keep fingers away  from probe contacts. Keep 
fingers behind the finger guards on the probes.

Measurements can be adversely affected by  impedances of additional 
operating circuits  connected in parallel or by transient currents.

Verify operation prior to measuring hazardous  voltages (voltages above 

30V AC RMS, 42V AC peak and 60V DC).

Do not use the meter with any parts removed.

Do not use the meter around explosive gas, vapor or dust.

Do not use the meter in a wet environment.

Specifications

Phase Rotation 

Clockwise or Anti-clock wise

Visual Indication 

LCD

Operating Temperature 

0°C to 40°C

Humidity 

 

15 to 80%

Operating Altitude 

2000m

Pollution Degree 

2

Type of Protection 

IP40

Battery 

 

9V

Current Consumption 

Max 20mA

Dimensions  

130 x 69 x 32mm