Operation Manual easyTymp
8111254 Rev. 7.0
41
02/16
Figure 53
3.
Insert the blue end of the floss from back to
front through one of the probe channels. Pull the
floss along its entire length through the channel
(Figure 53).
Figure 54
4.
Proceed in the same way with all 4 probe
channels. Use the floss only once (Figure 54).
Figure 55
5.
Place the probe tip back onto the probe.
Make sure that the plastic pegs are inserted into
the appropriate corresponding cavities (Figure 55).
Figure 56
6.
Screw the probe cap back on the probe
(Figure 56). The force of tightening the cap will
tighten the screw sufficiently. Never use tools to fix
the probe cap!
If any blockage or damage occurs to the sealing
gasket, the probe system can only be serviced by
MAICO.
Summary of Contents for easyTymp
Page 1: ...Operation Manual easyTympTM...
Page 2: ......