49
The
µMaxµm II
Series indicator has the ability to output one reading at a time or continu-
ous readings all the time. It can also have the serial number as part of the data output with
each specific reading that is sent.
When the indicator is set for single data output the data is triggered manually by the
M/OFF Data button.
When the indicator is set for continuous data output no manual trigger is needed.
When the data output is set to include the indicators serial number (S/N) the
S/N will be sent first then the reading delimited by a comma.
The Data Stream for the
µMaxµm II
is as follows:
NNNNNNN,±(D)DD.DDDD(D)_inch<CR>
There will be only one leading zero
The serial number identifies the unique indicator on the production line. It also can be
monitored for calibration or certification service requirements by use of a system software
program.
_
Represents Spaces
<CR>
Represents Carriage Return
N
Represents Serial Number Digit
D
Represents Digit on Display
inch
Units represented by inch or mm
(D)
Represents a possible digit depending on preset
and resolution