32
Figure 23 -
Continuous Data Output Mode displays new setting
The selection is toggled using the ZERO/ON button.
NOTE:
The continuous output operation may inhibit use of certain data collection cables,
devices, or software.
5.11 Calibration Mode (M123) [S]
1) Fit the indicator into a calibration device so that it is “on-scale”. The indicator
should remain “on-scale” throughout calibration. Select the inch or metric units
while in the gaging/measurement mode.
2) To enter Calibration Mode from the Gaging/Measurement Mode:
a) Simultaneously press and hold
M/OFF
(1), then quickly press and release
inch/mm
(2) once to enable the setup mode. (
M1
icon appears in the
lower left corner of the LCD) (Figure 3).
b) Press and release
M/OFF
to enter
M2
, Preset Mode.
c) Again, press and release
M/OFF
to enter M12, Limit #1 Tolerance Entry
Mode.
d) Again, press and release
M/OFF
to enter M23, Limit #2 Tolerance Entry
Mode.
e) Again press and release
M/OFF
to enter M13, Resolution Mode.
f) Press and hold
M/OFF
for 2+ seconds until M3 is displayed.
g) Press and release
M/OFF
to enter the M1 X1.
h) Press and release
M/OFF
to enter the M2 X1.
i) Press and release
M/OFF
to enter the M3 X1.