50
Mahr GmbH • Millimar C1202
5.1.5 Number of masters
–
Select the
Number of masters
submenu using the
▼
or
▲
key.
–
Select using
4
key or exit submenu
using the
Ó
key.
–
Select the number of masters using
the
▼
key.
1 = one-point master measurement
2 = two-point master measurement
see
Section 3.2.2. Master
–
Use the
4
key to confirm the
selection or
Ó
to exit the submenu.
–
Use the key to switch to
measuring mode.
5.1.6 Master reset
–
Select the
Master reset
submenu
using the
▼
or
▲
key, or exit the
submenu using the
Ó
key
or switch to measuring mode using
the key.
–
Use the
4
key to delete the master
values set in measuring mode. C
1202 switches to measuring mode.
The values displayed now refer
directly to the probe signal.
Master value is set
Master value is
deleted