3.2.2 Stationary Measuring Station
During stationary operation, the MarSurf PS1 can
be attached to the measuring stand (see Fig. 5).
Measuring stands of types STD, STF and STG
can be used.
A mount (6910201) is required for this which is
available as an optional accessory.
To fit the MarSurf PS1 on the measuring stand:
1. Set up the drive unit (see Section 3.2.1).
The pickup protection is not absolutely es
sential for operation on a measuring stand.
2. Lay the MarSurf PS1 on its top surface and
put the mount (22) on top of it so that the
mount’s two screws fit into the threaded
inserts on the bottom of the MarSurf PS1.
3. Tighten the two screws on the mount.
4. Introduce the mount’s bolt into the orifice on
the cage of the column and clamp it in place
using the two set screws (on the reverse of
the cage).
5. Align the testpiece and the drive unit using
the adjusting screw (21) so that they are
parallel.
The MarSurf PS1 can be inclined by ± 15°.
6. Rotate the handwheel for height adjustment
(20) to lower the pickup to the level of the
testpiece.
7. To change the testpiece, the MarSurf PS1 can
be tilted upwards in the mount and locked in
place.
It is essential that the two screws on the
mount and the two set screws are prop
erly tightened as otherwise the MarSurf
PS1 could come off and be damaged!
Fig. 5
Stationary measuring station
with measuring stand
20 Handwheel for height adjustment
21 Adjusting screw to correct inclination
22 Mount for MarSurf PS1 (6910201)
20
21
22