Mahr GmbH, MarSurf PS1
3.2
Setting up the Measuring Station
Practical examples:
– MarSurf PS1 standing on its veeblock bot
tom or on the vertical adjusters, measure
ment on level surface or in bore, stylus tip
pointing downwards.
– MarSurf PS1 lying upsidedown on its top
surface (threepoint support), veeblock
serving as mount, measurement on cylindri
cal testpieces of up to 1 kg
(2.205 lbs)
, stylus
tip pointing upwards.
Depending on the length of the testpiece,
the drive unit can be moved longitudinally
into the front or rear lockin position.
3.2.1 Mobile Measuring Station
To set up the drive unit:
1. Carefully introduce the pickup (5) into
the pickup holder on the end face of the
drive unit (7), bearing in mind the plug
assignment.
2. Screw the pickup protection (6) firmly onto
the end face of the drive unit.
The pickup protection should be used
at all times if possible as it protects
the pickup from damage.
Exception: When measuring in
smalldiameter bores or at points
that are difficult to access.
After a pickup has been replaced,
the new pickup should be calibrat
ed or its correction value set (if this
is already known) (see Section 5).
3. Fit any additional accessories, e.g. for vertical
adjustment – i.e. the mount (8) and one or
both vertical adjusters (9).
4. Place the drive unit in the correct measuring
position by moving it longitudinally and
rotating it. Guide pins on the drive unit help
to find the ideal operating position.
Summary of Contents for MarSurf PS1
Page 2: ......
Page 3: ...Mahr GmbH MarSurf PS1 Deutsch English Français Italiano Español ...
Page 4: ...Mahr GmbH MarSurf PS1 ...