Mahr GmbH • Millimar C1202
The instrument is most accurate after approx. 30 minutes of operation. To reliably
balance two probes, the probes’ electrical zero points should be positioned at the
calibration zero points using the probe raw value display. (see point 6.1.1.).
This is because the inductive probe measures most accurately at the electrical zero
point. A 1.000 mm gage block
would be the perfect electrical zero point in our example, where a 2.000 mm gage
block is used for the MAX value and the measuring plate represents the MIN value.
6.1.2.3 Symmetry C1 to C2
When balancing two probes, a customer calibration is performed on the two probes
at both measuring channels in succession.
–
Select the
Calibration
submenu
using the
▼
or
▲
key.
–
Select using
4
.
–
Select the
Symmetry C1 to C2
submenu using the
▼
or
▲
key.
–
Select using
4
key or exit submenu
using the
Ó
key.
– Use
▼
to select the value for
Calibration min
.
–
Use the
4
key to select the
Calibration min
submenu.
–
Use the
▼
key to select the digit and