Magtrol Model 6400 Torque Transducer Display
Chapter 4 – Manually Controlled Operation
OPERA
TION
Torque Scale Factor Setup
Sets the full scale rating of the torque transducer being used. Select from a range of 0.0001 N / 5
volts to 100000 N / 5 volts.
1. Press and release SHIFT. The display appears as follows:
POWER/AUX
TORQUE
SPEED
Figure 4–7 Torque Scale Factor Setup Display
2. Use UP
S
, DOWN
T
, LEFT
W
and RIGHT
X
buttons to set torque scale factor (Q
SCALE).
Max Power Setup
Sets the maximum power set point. Select from a range of 0.1 watts to 100,000 kilowatts.
1. Press and release SHIFT. The Max Power Setup Display (as seen in Figure 4–3) appears.
2. Use UP
S
, DOWN
T
, LEFT
W
and RIGHT
X
buttons to select the maximum power set
point. The maximum power set point is used to indicate an overload condition. If the
power reading exceeds what was programmed, an OVERLOAD WATTS message appears
in the display (as seen in Figure 4–4 Power Overload Display).
Direction Input Setup
When the direction input function is enabled, it tells the 6400 to read the rotational sense bit from
the torque transducer and displays that information in the SPEED portion of the display. A positive
(+) reading indicates counterclockwise rotation while a negative (-) reading indicates clockwise
rotation.
1. Press and release SHIFT. The Direction Input Setup Menu (as seen in Figure 4–5) appears.
2. Use UP
S
and DOWN
T
buttons to turn direction input ON or OFF.
3. Press and release SHIFT to return to main menu.
4.1.4
A
UXILIARY
S
CALE
F
ACTOR
S
ETUP
Sets the scale factor for auxiliary input. Select from a range of 0.0001 to 1000 units per volt.
Default is 1 unit/volt.
1. Press and release SHIFT.
2. Press and release AUX SETUP button. The display appears as follows:
POWER/AUX
TORQUE
SPEED
Figure 4–8 Auxiliary Scale Factor Setup Display