Magtrol Model 6200 Open Loop Dynamometer Controller
Chapter 5 – Computer Controlled Operation
OPERA
TION
D5
D6
D7
D8
REN
DAV-COM
NFRD-COM
NDAC-COM
IFC-COM
SRQ-COM
ATN-COM
SIGNAL GROUND
IEEE-488 INTERFACE
D1
D2
D3
D4
EO1
DAV
NFRD
NDAC
IFC
SRQ
ATN
SHIELD
1 13
2 14
3 15
4 16
5 17
6 18
7 19
8 20
9 21
10 22
11 23
12 24
Figure 5–1 GPIB (IEEE-488) Interface
5.1.2
C
Hanging
tHe
gPiB P
rimary
a
ddress
Each instrument serviced by the gPIb has its own Primary Address code, which enables the
computer to obtain readings from the instrument. The factory default setting on the 6200 is 15.
Some PC interfaces can access from one to fifteen 4-bit primary addresses. other interfaces can
access as many as thirty-one 5-bit primary addresses. The 6200 uses the 4-bit format.
1. Press the ShIFT button and release.
2. Press SETUP.
3. Press DoWN
twice.
4. Press ShIFT twice.
5. Press
or DoWN
until you see the desired gPIb address.
6. Press ShIFT twice to exit.
5.2
CheCking the 6200-to-pC ConneCtion
Note:
Make sure that the 6200 and its host computer are communicating
before acquiring data.
1. Make sure the primary address is set correctly for the 6200.
2. Set the input variable to 15 characters (13 variable characters and the two required data
termination characters CR and LF. (See “Programming” in this chapter.)
3. Issue output data command “OD” and read 15 characters according to the instructions for
your gPIb interface.