MAGSTIM BiStim2 Operating Manual Download Page 7

 

Magstim BiStim² 

 

 

Operating Manual 

 

© The Magstim Company Limited 

MOP02-EN-02 

 

 

 

 

Explosive  and  Flammable  Atmospheres: 

The  Magstim  Bistim²  and  accessories  must  not  be 

used  in  an  explosive  atmosphere,  around  explosive  gases  or  in  the  presence  of  flammable 
anaesthetics. 

 

Protective Earth: 

To avoid the risk of electric shock, the Magstim Bistim² must only be connected 

to a supply mains with protective earth.

 

 

Modification: 

No modification of this equipment is allowed.

 

 

Discharge  noise:

  when  the  magnetic  pulse  is  delivered,  a  discharge  click  is  produced  that  may 

startle. The use of ear protection is recommended. 

 

Seizures:

  cortical  magnetic  stimulation  runs  the  risk  of  inducing  seizures.  Magstim  concurs  with 

the  advice  given  in  the  paper  titled  “Safety,  ethical  considerations  and  application  guidelines  for 
the  use  of  TMS  in  clinical  practice  and  research”  by  Rossi  et  al  (2009)  in  respect  to  screening 
candidates prior to administering TMS. 

 

Environmental  Conditions

*

The  system  must  not  be  used  or  stored  under  environmental 

conditions that fall outside those specified in section 6.3 of this operating manual. 

 

Stacking  Limit: 

Magstim  200²  units  may  be  stacked  on  top  of  each  other,  but  must  not  exceed 

the stacking limit specified in Section 6 of this operating manual. If this limit is exceeded, the stack 
may  become  unstable  and  overbalance  risking  injury  to patients  or  users  and  risking  damage  to 
the unit’s themselves. 

If  the  stacking  limit  is  reached  it  can  have  a  weight of  up  to  80kg. Ensure  that the  surface  upon 
which to system is to be placed is capable of supporting this weight. Failure to do so may result in  
injury or damage. The weight for each individual 200² unit is in excess of 20kg

 

 

Carrying

*

On  no  account  should  the  Bistim²  connecting  module  be  used  as  a  carrying/gripping 

point when transporting its host Magstim 200² unit. The Bistim² connecting module and its fixings 
are  not  designed  to  be  subjected  to  this  level  of  force  and  the  Bistim²  connecting  module  will 
become  detached  from  its  host  unit.  This  could  result  in  damage  to  Magstim  200²  unit,  Bistim² 
connecting module, or those transporting the equipment

 

 

Connecting Equipment: 

Only equipment listed in Section 1.3 may be connected to the Magstim 

BiStim².  Equipment that meets the relevant IEC standard may be connected to the rear  interface 
ports at the rear of the Magstim Bistim². See Appendix B for further details.

 

Summary of Contents for BiStim2

Page 1: ...January 2014 Magstim BiStim Operating Manual MOP02 EN Revision 02...

Page 2: ...4 4 Operation 14 4 5 Independent BiStim Triggering IBT 15 Section 5 Maintenance 16 5 1 User Maintenance and Calibration 16 5 2 Technical Maintenance 16 5 3 Voltage Selection and Fuse Rating 17 5 4 Cle...

Page 3: ...al service Claims for damages during shipment must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as th...

Page 4: ...gstim Bistim is a non sterile reusable medical device intended for multi patient use The Magstim Bistim is not intended to come into contact with the patient as this achieved through use of a stimulat...

Page 5: ...ent environment Note When the Bistim Connecting Module is not connected the Magstim 200 Units will operate as if they were a standalone device Please consult the Magstim 200 Operating Manual before us...

Page 6: ...d to the Magstim Bistim This connection must be configured in compliance with IEC 60601 1 1 2000 or Clause 16 of IEC 60601 1 2005 Second edition IEC 60601 1 1 with the following interface voltage limi...

Page 7: ...ng manual Stacking Limit Magstim 200 units may be stacked on top of each other but must not exceed the stacking limit specified in Section 6 of this operating manual If this limit is exceeded the stac...

Page 8: ...trol system The only function button that may be used is the system stop button Figure 3 1 Detailed front panel The upper unit is considered the master unit The lower unit is considered the slave unit...

Page 9: ...g connected correctly to the Magstim Bistim 3 1 6 Coil Temperature Indicators 6 These LED s represent the surface temperature of the coil The LED s illuminate sequentially as the coil temperature rise...

Page 10: ...eases the level To toggle between the master power slave power and inter pulse spacing press the output control Each complete 360 rotation changes the master slave power output level by approximately...

Page 11: ...MOP02 EN 02 Note It is normal for the pin indicated on the label to protrude from the coil socket on the master and slave stimulator units This is an intentional design feature Figure 3 2 Coil output...

Page 12: ...xternal trigger input This is only functional on the master unit Please see Appendix B for more information 3 2 3 Foot Switch Socket 3 This socket is used to connect the pneumatic foot switch to the M...

Page 13: ...unit prior to the unit being switched on 3 3 Accessories A range of stimulating coils are available for use with the Magstim range of products For more details please contact the Sales department Cont...

Page 14: ...ating Manual Note The contacts in the coil connector Bistim connecting module and the stimulator coil output socket should be checked regularly for any signs of tarnishing or burning See Section 5 2 f...

Page 15: ...key tighten the bolts so as to secure the coil holder to the right hand side of the master unit Do not over tighten the bolts as this may damage the cases of the units Figure 4 2 Attaching the coil ho...

Page 16: ...l holder on the right hand side of the stimulator as instructed in Figure 4 3 Figure 4 4 Attaching the coil to the coil holder 5 Connect the stimulating coil to be used to the coil output socket on th...

Page 17: ...ster unit UI to adjust the power level or the inter pulse spacing to the desired setting Rotating the control clockwise increases the power level rotating anti clockwise decreases the power level 4 De...

Page 18: ...nnect the Magstim Bistim Note Ensure that the system is disarmed prior to disconnecting the stimulating coil and BiStim connecting module 4 5 Independent BiStim Triggering IBT When the system is set i...

Page 19: ...ossible for localised heating to manifest itself in the connector which causes damage to the contacts Continued use in this condition will eventually cause the contacts to deteriorate and loose electr...

Page 20: ...voltage selector to the desired setting and replace the cover A legend will indicate the voltage selected when the fuse cover is replaced Note Only replace fuses with those of the correct rating and...

Page 21: ...erviceable life it should not be disposed of in general waste The Magstim Company Limited should be contacted see Section 7 2 for contact details for advice on its disposal in compliance with the appr...

Page 22: ...ot comply with EMC Safety Standard EN 60601 1 2 including mobile phones Any interface cable which connects the Magstim Bistim unit to an external piece of equipment must be no more than 1 5m in length...

Page 23: ...ife expectancy 2 x 10 6 discharges at 70 power level 8 x 10 5 discharges at 80 power level 4 x 10 5 discharges at 90 power level 2 x 10 5 discharges at 100 power level Isolated Interface Logic High Vo...

Page 24: ...ering 6 5 Packing Instructions If for any reason it is necessary to return your Magstim Bistim care should be taken to ensure that the equipment is adequately packed to prevent transit damage Ideally...

Page 25: ...imited Spring Gardens Whitland Carmarthenshire SA34 0HR Telephone 44 0 1994 240798 Fax 44 0 1994 240061 E mail info magstim com Website www magstim com 7 2 Servicing Enquiries Telephone 44 0 1994 2429...

Page 26: ...gory C04 Invalid Power Identification C05 Temperature sensor circuit failure C06 Bad Serial command or received data checksum C07 Internal Software Error C08 Invalid Coil Temperature ID C09 Invalid te...

Page 27: ...ck max difference E75 HVCAP over temperature E76 HV Transformer over temperature E77 Charge threshold failure E78 VREF check failure E79 HVCAP voltage comparison failure E80 Charging fault E81 HV over...

Page 28: ...Bistim will only be triggered when the logic level being applied on the Tin or Tin pin is changed On the Tin pin triggering will occur when the logic level is changed from low to high On the Tin pin...

Page 29: ...romagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Magstim Bistim can help prevent electromagnetic interference by maintaining a minimum distance b...

Page 30: ...ital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT f...

Page 31: ...tance in metres m Field Strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...