MAGSTIM BiStim2 Operating Manual Download Page 13

 

Magstim BiStim² 

 

 

Operating Manual 

 

© The Magstim Company Limited 

10 

MOP02-EN-02 

3.2.4 

Cooling Fan Outlet (4) 

With the  Magstim  Bistim²  turned  on,  air  is  drawn  through  slots  at the  bottom  of the  instrument  and 
expelled through the fan outlet. There should be no obstructions placed closer than 50mm to the fan 
outlet. 

 

3.2.5 

Functional Earth (5) 

Functional  earth  connections  of  the  master  and  slave  units  should  be  connected  using  the  supplied 
functional earth strap. 

 

3.2.6 

Power Entry Module & Fuse Access (6) 

The standard IEC mains lead supplied by the Magstim Company Ltd should be connected to the unit 
prior to the unit being switched on.  

 

3.3 

Accessories 

A  range  of  stimulating  coils  are  available  for  use  with  the  Magstim  range  of  products.  For  more 
details, please contact the Sales department. Contact details are shown in Section 7.3. 

Please note, only “category A” high power stimulating coils will work with the Magstim Bistim². 

Note:

 The use of the correct accessories is essential  to the functioning of the  Magstim Bistim². The 

Magstim Company Limited 

cannot guarantee the instrument’s performance unless accessories used 

are  obtained  from  The  Magstim  Company  Limited.  Only  Magstim  accessories  purchased  from  the 
The Magstim Company Limited should be used with the Magstim Bistim². 

Note: 

The Magstim BiStim² should  only be powered  using the power lead supplied by The Magstim 

Company Ltd. 

 

 

 

Summary of Contents for BiStim2

Page 1: ...January 2014 Magstim BiStim Operating Manual MOP02 EN Revision 02...

Page 2: ...4 4 Operation 14 4 5 Independent BiStim Triggering IBT 15 Section 5 Maintenance 16 5 1 User Maintenance and Calibration 16 5 2 Technical Maintenance 16 5 3 Voltage Selection and Fuse Rating 17 5 4 Cle...

Page 3: ...al service Claims for damages during shipment must be filed promptly with the transportation company All correspondence concerning the equipment must specify the model name and or number as well as th...

Page 4: ...gstim Bistim is a non sterile reusable medical device intended for multi patient use The Magstim Bistim is not intended to come into contact with the patient as this achieved through use of a stimulat...

Page 5: ...ent environment Note When the Bistim Connecting Module is not connected the Magstim 200 Units will operate as if they were a standalone device Please consult the Magstim 200 Operating Manual before us...

Page 6: ...d to the Magstim Bistim This connection must be configured in compliance with IEC 60601 1 1 2000 or Clause 16 of IEC 60601 1 2005 Second edition IEC 60601 1 1 with the following interface voltage limi...

Page 7: ...ng manual Stacking Limit Magstim 200 units may be stacked on top of each other but must not exceed the stacking limit specified in Section 6 of this operating manual If this limit is exceeded the stac...

Page 8: ...trol system The only function button that may be used is the system stop button Figure 3 1 Detailed front panel The upper unit is considered the master unit The lower unit is considered the slave unit...

Page 9: ...g connected correctly to the Magstim Bistim 3 1 6 Coil Temperature Indicators 6 These LED s represent the surface temperature of the coil The LED s illuminate sequentially as the coil temperature rise...

Page 10: ...eases the level To toggle between the master power slave power and inter pulse spacing press the output control Each complete 360 rotation changes the master slave power output level by approximately...

Page 11: ...MOP02 EN 02 Note It is normal for the pin indicated on the label to protrude from the coil socket on the master and slave stimulator units This is an intentional design feature Figure 3 2 Coil output...

Page 12: ...xternal trigger input This is only functional on the master unit Please see Appendix B for more information 3 2 3 Foot Switch Socket 3 This socket is used to connect the pneumatic foot switch to the M...

Page 13: ...unit prior to the unit being switched on 3 3 Accessories A range of stimulating coils are available for use with the Magstim range of products For more details please contact the Sales department Cont...

Page 14: ...ating Manual Note The contacts in the coil connector Bistim connecting module and the stimulator coil output socket should be checked regularly for any signs of tarnishing or burning See Section 5 2 f...

Page 15: ...key tighten the bolts so as to secure the coil holder to the right hand side of the master unit Do not over tighten the bolts as this may damage the cases of the units Figure 4 2 Attaching the coil ho...

Page 16: ...l holder on the right hand side of the stimulator as instructed in Figure 4 3 Figure 4 4 Attaching the coil to the coil holder 5 Connect the stimulating coil to be used to the coil output socket on th...

Page 17: ...ster unit UI to adjust the power level or the inter pulse spacing to the desired setting Rotating the control clockwise increases the power level rotating anti clockwise decreases the power level 4 De...

Page 18: ...nnect the Magstim Bistim Note Ensure that the system is disarmed prior to disconnecting the stimulating coil and BiStim connecting module 4 5 Independent BiStim Triggering IBT When the system is set i...

Page 19: ...ossible for localised heating to manifest itself in the connector which causes damage to the contacts Continued use in this condition will eventually cause the contacts to deteriorate and loose electr...

Page 20: ...voltage selector to the desired setting and replace the cover A legend will indicate the voltage selected when the fuse cover is replaced Note Only replace fuses with those of the correct rating and...

Page 21: ...erviceable life it should not be disposed of in general waste The Magstim Company Limited should be contacted see Section 7 2 for contact details for advice on its disposal in compliance with the appr...

Page 22: ...ot comply with EMC Safety Standard EN 60601 1 2 including mobile phones Any interface cable which connects the Magstim Bistim unit to an external piece of equipment must be no more than 1 5m in length...

Page 23: ...ife expectancy 2 x 10 6 discharges at 70 power level 8 x 10 5 discharges at 80 power level 4 x 10 5 discharges at 90 power level 2 x 10 5 discharges at 100 power level Isolated Interface Logic High Vo...

Page 24: ...ering 6 5 Packing Instructions If for any reason it is necessary to return your Magstim Bistim care should be taken to ensure that the equipment is adequately packed to prevent transit damage Ideally...

Page 25: ...imited Spring Gardens Whitland Carmarthenshire SA34 0HR Telephone 44 0 1994 240798 Fax 44 0 1994 240061 E mail info magstim com Website www magstim com 7 2 Servicing Enquiries Telephone 44 0 1994 2429...

Page 26: ...gory C04 Invalid Power Identification C05 Temperature sensor circuit failure C06 Bad Serial command or received data checksum C07 Internal Software Error C08 Invalid Coil Temperature ID C09 Invalid te...

Page 27: ...ck max difference E75 HVCAP over temperature E76 HV Transformer over temperature E77 Charge threshold failure E78 VREF check failure E79 HVCAP voltage comparison failure E80 Charging fault E81 HV over...

Page 28: ...Bistim will only be triggered when the logic level being applied on the Tin or Tin pin is changed On the Tin pin triggering will occur when the logic level is changed from low to high On the Tin pin...

Page 29: ...romagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Magstim Bistim can help prevent electromagnetic interference by maintaining a minimum distance b...

Page 30: ...ital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT f...

Page 31: ...tance in metres m Field Strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur...

Reviews: