display
Password
action
Comment
13
SZHeight
xx.x su
User
Enter
distance below BlockDis where
SZ Fault will be asserted
5.1 to 2286 cm (2 to 900 in)
14
SZ Latch
Reset
User
Press
Enter to clear a Safety Zone latch
15
Threshld
(select)
User
Select
from CFD, Fixed
For interface, refers to threshold for
upper level pulse
16
Trim Lvl
xx.x su
Superuser or
user
Enter
value to adjust Level reading
-20.0 inches <= Lvl Trim <= +20.0 inches
(Superuser password required if negative
fiducial.)
17
AI Disp
User
Transmitter Display
AI Block Display
18
LvlTicks
xxxxx
User
Diagnostic Display
Time of flight from fiducial to level signal
19
New Pass
xxx
User
Enter
new password (0-255)
Displays encrypted value of present
password
20
Language
(select)
User
Select
from English, Spanish, French,
German
Language choice for
LCD display
21
Mdl705FF
Ver 3.xx
None
Transmitter Display
Product identification
Firmware version
22
DispFact
(select)
None
Select
Yes to display factory parameter
menus
23
History Status
None
Diagnostic Display to view present status
and recent exceptions
24
Run Time
xxxx.x h
None
Diagnostic Display showing elapsed
time since power on
Cleared to zero with History Reset
25
History
Reset
Superuser
Press
Enter and select yes to clear
history
Reset History Data
26
HF Cable
Select
Superuser
Select length of remote
extension cable
Select from integral, 3-foot, 12-foot
27
FidTicks
xxxx
None
Diagnostic Display
Time of flight from start of
ramp to fiducial
28
Fid Sprd
xxx
None
Diagnostic Display
Spread in fiducial ticks readings
29
Fid Type
(select)
Superuser
Superuser parameter
Select
from Positive, Negative.
Selection only allowed for some probes,
fixed for others
30
Fid Gain
xxx
Superuser
Superuser parameter
31
Window
xxx
Factory
Factory Parameter
Calibration parameter
32
Conv Fct
xxxx
Factory
Factory Parameter
Calibration parameter
33
Scl Ofst
xxx
Factory
Factory Parameter
Calibration parameter
36
57-640 Eclipse Guided Wave Radar Transmitter - F
OUNDATION
fieldbus
™