LH72 initial Setup Manual
・
Display resolution is set by pressing
key on each axis.
・
A value smaller than the input resolution cannot be selected.
・
Polarity changes with
key
・
To display the diameter (double display), turn on on the
display axis.
Scaling magnification input
≤
≤
magnification
Diameter display
Display resolution
SCALING
(scaling)
Angle display
Length display
Extended resolution
Press the edit key
How to set up Detailed Settings (1/1)
(continued from Basic Settings)
Step 4
Make the detailed settings.
*If you have already completed the basic settings and want to
reconfigure only the advanced settings, you can do the same.
Basic Settings
Press the setup key
in the normal count display state.
D
e
ta
ile
d
S
et
tin
gs
default
No compensation
Error compensation off
Linear compensation
Linear Error
Segmented error compensation
Segmentation Error
Error Setting
(compensation value)
Selection of error compensation function
default
~
Input a correction value of
±
0.600 mm for a distance
of 1 m. Input up to
±
1.000 mm when extended is
selected.
~
Extended res.
Normal
Enter with the number keys
Segmentation Error
(
Segmented error compensation
)
For the operation method, refer to the instruction
manual or the simple manual (Segmented error
compensation).
Segmented error compensation function uses a scale
with reference point to correct mechanical errors for
each interval.
The range that can be corrected is the range of the
counter display based on the reference point position.
In order to use this function, it is necessary to
measure the section error of the machine with a laser
interferometer, length reference device, etc. in
advance.
Liner Error
(Linear compensation)
default