LH71 Initial setup (Example)

Ver.1 (2021.1) CS&S

When the display is                             , 

press the             key once.

The display will be                              . (Destination selection mode)

Press           key on the right side once.

The display will be                            . 

( If you want to change the destination, press             key. )

When you have made your selection, press            to save your settings. 

TYPE

Measuring unit
SJ700A (3 axes)
Output resolution: 5μm

SET UP

After about 
2 seconds

For LH71A-3

Display unit: LH71-3 (3 axes input)

Z axis

X axis

Y axis

SJ700A

SJ700A

SJ700A

Press              key once.

The display will be                          . (Measurement unit resolution)

Press          key to check the current measurement unit resolution.

If the resolution is not 5μm, press           key to change it.

After checking the resolution, press            key to save the settings.

Switch from basic settings to normal count mode.

Press 

key once and the display will change to 

.

Press 

key once and the display will change to

.

Press 

key once to automatically switch to advanced settings mode.

The display will be 

. Press 

key once.

Confirm that the display is 

display.

(Press 

key to change the polarity)

When the selection is complete, press 

key to save the settings.

Connect the AC adapter and
Turn on the primary power supply

(Not            key)

When you press 

key, the display becomes the normal display.

This completes the initial settings

AC adapter is energized.

Turn on 

key, 

the LED light changes 

from “blinking” to “off” and then 

Press and hold 

key 

for 3 seconds.

The following display appears

When the power is turned 

on for the first time after 

shipment from the factory

When you want to change 

the basic settings

START

Set to basic setting mode

Finished

AC primary power source

Setting conditions

Setting items

Setting value (selection)

Basic 

settings

TYPE

GENERAL (General machine )

ADD(Addition display)

Not displayed when GENERAL is selected

Destination

STD

Measuring unit 
resolution

Detailed 
settings 

Display resolution* 
and polarity

Display switching

Input 1: Upper display

Input 2: Middle display
Input 3: Lower display

Axis label

Upper : X, Middle : Y, Lower : Z

Scaling

1 time

1.000000)

Compensation value

No setting

OFF

Flicker control

Strong

2

Sleep

No setting

OFF

The detailed settings are the factory default settings.

* The initial value is the measurement unit resolution set in the basic settings.

default

default

default

default

default

default

default

Input 1

Input 2

Input 3

Confirm that the resolution of 

input 1,2, and 3 is 5 μm (SJ700A scale).

* 10 μm “10U” for Digiruler PL20C