![Magic Care MIZAR Instructions For Use Manual Download Page 20](http://html2.mh-extra.com/html/magic-care/mizar/mizar_instructions-for-use-manual_3433718020.webp)
20
Mod. P0915EM
Power supply:
12V DC 1,25A
Max pressure:
2 ± 0.3 bar
Compressor air delivery:
approx 9 l/min
Noise (at 1 m):
approx 51 dB (A)
Operation use:
30 minutes ON/ 30 minutes OFF
Operating Conditions:
Temperature:
min 10°C; max 40°C
Air humidity:
min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Storage conditions:
Temperature:
min -25°C; max 70°C
Air humidity:
min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Dimensions: 12 (W) x 12 (D) x 6 (H) cm
Weight:
0.58 Kg
Power supply
Only use power supply mod. UME318-1215 (cod. 17111)
Primary: 100/240V~ 50/60Hz
Secondary: 12V DC 1.5 A
APPLIED PARTS
Type BF applied parts are:
patient accessories (C2, C3, C4, C5, C6.5)
RF8 Dual Speed nebuliser
Medication minimum capacity: 2 ml
Medication maximum capacity: 8 ml
Operating pressure (with neb.):
0.8 bar
TECHNICAL SPECIFICATIONS
C1.2.3 speed selector
in pos Max
in pos Min
(1)
Delivery:
0.40 ml/min approx.
0.19 ml/min approx.
(2)
MMAD:
4.48 μm
4.01 μm
(2)
Breathable fraction < 5 μm (FPF):
55.5 %
61.1 %
(1) Data recorded according to internal Flaem procedure I29-P07.5.
(2) In vitro characterization certified by TÜV Rheinland LGA Products GmbH - Germany in compliance with
European Standard for Aerosol therapy apparatuses EN 13544-1. Further details are available on request.
Rhino Clear® Ergo NASAL DOUCHE
Maximum medication capacity:
10 ml
Delivery Time 10 ml(3):
2 minutes
(3)
Tests performed with 0.9% NaCl saline solution, according to the Flaem internal procedure.