![Magic Care MISTRAL PLUS P0106EM F400 Instructions For Use Manual Download Page 12](http://html1.mh-extra.com/html/magic-care/mistral-plus-p0106em-f400/mistral-plus-p0106em-f400_instructions-for-use-manual_3432668012.webp)
12
13
the methods described below.
method A: Accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of
potable water and disinfectant, according to the proportions indicated on the
packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of
air bubbles on the parts. Leave the parts immersed for the amount of time indicated
on the packaging of the disinfectant associated with the concentration chosen for
the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Disinfect the accessories C1.1-C1.2-C1.3-C2-C3 by boiling them in water
for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Disinfect the accessories C1.1-C1.2-C1.3-C2-C3 with a hot steam steriliser
for baby-bottle (not the microwave type). Perform the process faithfully following
the instructions of the steriliser. To ensure that the disinfection is effective, choose a
steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on
a paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair
dryer).
REPLACING THE FILTER
The device has a suction filter (A3) that should be replaced
when it is dirty or changes color. Do not wash or reuse the
same filter. Regular replacement of the filter is required to
help and ensure correct performance of the compressor.
Check the filter on a regular basis. Contact your local dealer
or authorized service center for spare filters.
To replace the filter take it out by pulling it as shown in figure.
The filter has been designed so that it is always fixed in its seat.
Use original FLAEM accessories only.