SOFTTOUCH FACE MASKS
Soft
biocompatible
material
Dispersion
Limiting Vent
CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from the
socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of
any sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.2) anticlockwise, remove the upper
part (C1.2 and C1.3) as shown in the “Assembly diagram” in point C1.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one
of the methods described below.
method A: Sanitise accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 under potable hot
water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 in the dishwasher with
a hot cycle.
method C: Sanitise accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 by immersing them in
a solution of 50% water and 50% white vinegar, then rinse thoroughly under potable
hot water (approximately 40°C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of
SoftTouch masks have an outer edge made of soft
biocompatible material that ensures excellent
adherence to the face, and is also equipped with
an innovative Dispersion Limiting Device. These
distinctive elements that distinguish it allow great-
er sedimentation of medication in the patient and
also limit dispersion.
During the inspiratory phase,
the tab, acting as a Dispersion
Limiting Device, bends inwards
towards the mask.
During the expiratory phase, the
tab, acting as a Dispersion Lim-
iting Device, bends outwards
from the mask.
deeply in and out. After inhaling, we recommend holding your breath for an instant
to allow the inhaled drops of aerosol to deposit. Then exhale slowly.
6. Upon completing application, switch off the device and unplug it.
ATTENTION: If after the therapy session an evident deposit of moisture forms within
the pipe (B), detach the pipe from the nebuliser and dry it with the very ventilation
from the compressor; this operation prevents possible blooms of mould inside the
pipe.