18
15. Time Adjustment
1.
To adjust the date/ time in the monitor, press
the
Key.
The display will show a blinking number
showing the month.
2.
Change the month by pressing the
Key.
Each press will increase the number by one in
a cycling manner. Press the
Key
again to
confirm the entry, and the screen will show a
blinking number representing the date.
3.
Change the date, the hour, and the minute as
described in Step 2 above, using the
Key
to change and the
Key
to
confirm the entries.
4.
A
“
“0”
will reappear as the Blood Pressure
Monitor is ready for measurement again.
Press the
key
blinking month
increasing month
confirm the entry
16. Troubleshooting
If any abnormality will arise during use, please check the following points.
Symptoms
No display when
the
Key
is
pressed
EE mark shown on
display or the blood
pressure value is
displayed
excessively low
(high)
Check Points
Have the batteries run
down?
Have the batteries’
polarities been
positioned incorrectly?
Is the cuff placed
correctly?
Did you talk or move
during measurement?
Did you shake the
wrist with the cuff on?
Correction
Replace them with two new alkaline
batteries.
Re-insert the batteries in the correct
positions.
Wrap the cuff properly so that it is
positioned correctly.
Measure again quietly. Keep wrist
steady during measurement.
Note:
If the unit still does not work, return it to your dealer or an authorised service centre.
Under no circumstance should you disassemble and repair the unit by yourself.
To ensure accurate measurements, it is recommended that the performance should be
checked every 2 years.