Warning
•
The sound pressure level in the ear canal increases when you test patients with small ear canals.
Always comply with local practice and recommendations for presenting loud stimuli.
Whenever an intensity level exceeds the warning level (> 100 dB HL), a warning message will be shown, and you will be
prompted to decide whether to continue or to move on to the next stimulus type.
12.6
Reflex Threshold testing
Note
•
It is recommended that you perform a tympanometric test before making any acoustic reflex meas-
urement, and determine the acoustic reflex threshold before making a reflex decay measurement.
This test is supported by MADSEN Zodiac Diagnostic and MADSEN Zodiac Clinical.
You can record a reflex threshold either as a separate measurement or as part of a diagnostic sequence. In Sequence test-
ing, the test automatically suggests the threshold level. Always carefully review this result and adjust it if necessary.
Manual Reflex Threshold testing: see
Manual Reflex Threshold testing
Automated Reflex Threshold testing: see
Automated Reflex Threshold testing
.
Preparing for testing
1. Do as described in
.
2. Inform the patient about the high stimulus levels in the test.
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If contralateral stimulation is used in the test, make sure that both the ipsilateral probe and the contralateral phone
are in place before you start the measurement.
5. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
12.6.1
Manual Reflex Threshold testing
In the following you will find the description for performing manual reflex threshold testing as a single test. This means
that the automated threshold search is not enabled.
Otometrics - MADSEN Zodiac
93
12 Testing with Zodiac Diagnostic or Clinical - PC-based
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...