12.6.2
Automated Reflex Threshold testing
In the following you will find the description for performing automated reflex threshold testing. The automated search for
the reflex threshold can be used both for individual reflex stimulus types and for preconfigured sequence testing.
The following procedure is an example describing how you search for a single stimulus type threshold.
Starting the test
1. Select
Auto
testing.
2. Ask the patient to sit very still and quiet during the test, without moving
head or jaw.
3. Prepare the patient for the high sound levels in the test.
4. Click either the
Ipsi
reflex or the
Contra
reflex button.
5. Select the frequency that you wish to test: Click the applicable arrow but-
ton.
6. Check the baseline graph to make sure that you have a stable test con-
dition before you start presenting the stimulus. The baseline indicator
should be as close to zero as possible. If needed, you can offset the pres-
sure to stabilize your measurements: Click the applicable arrow button.
7. Make sure that the settings options for the automated reflex sequence
have been selected. They include starting level, increment size, veri-
fication procedure, and maximum level.
As opposed to a full measurement sequence, the automated threshold
search only includes a single, selected stimulus side and stimulus type.
Note
•
You will be warned about a high stimulus intensity level if you
reach the warning limit.
8. Click the
Start
button to start the automated threshold search. The deflec-
tion curve is plotted in real time.
9. If needed, repeat these steps until you have collected the desired meas-
urements for the stimulus sides and types you wish to test.
The baseline indicator starts imme-
diately when you insert the probe in
the ear of the patient.
Warning
•
If the patient is troubled by the test, stop the test. The test
is interrupted immediately. Already measured results are kept.
The search will be paused before each stimulus above 100 dB HL, and you will
be prompted to continue.
Stopping the test
The test stops automatically, but can be stopped manually, if needed.
1. Click the
Stop
button on the Control Panel or the
Start/Stop
button on the probe.
2. The test is interrupted immediately.
Otometrics - MADSEN Zodiac
95
12 Testing with Zodiac Diagnostic or Clinical - PC-based
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...