MadgeTech Therm-A-lert Product User Manual Download Page 9

9

Therm•A•lert

Troubleshooting

Why is the wireless data logger not appearing in the software?

If the data logger doesn’t appear in the Connected Devices panel, or an error message is  
received while using the data logger, try the following: 

 

»

Check that the RFC1000 is properly connected. For more information, see 

Troubleshooting 

wireless transceiver problems

 (below).

 

»

Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter 
connected to the battery of the device. If possible, try switching to a new battery. 

 

»

Ensure that the 

MadgeTech 4 Software

 is being used, and that no other MadgeTech Software 

(such as 

MadgeTech 2,

 or 

MadgeNET

) is open and running in the background. 

MadgeTech 2

 and 

MadgeNET

 are not the compatible with the RFRHTemp2000A.

 

»

Ensure that the 

Connected Devices 

panel is large enough to display devices. This can be verified 

by positioning the cursor on the edge of the 

Connected Devices

 panel until the resize cursor 

appears, then dragging the edge of the panel to resize it.

 

»

Ensure that the data logger and RFC1000 are on the same wireless channel. If the devices are 
not on the same channel, the devices will not communicate with one another. Please refer to the 
Channel Programming section for information on changing the device channel.

Troubleshooting wireless transceiver problems 

Check that the software properly recognizes the connected RFC1000 wireless transceiver.

If the wireless data logger is not appearing in the 

Connected Devices

 list, it may be that the RFC1000 

is not properly connected.

1. 

In the MadgeTech 4 Software, click the 

File

 button, then click 

Options

2. 

In the 

Options

 window, click 

Communications

3. 

The 

Detected Interfaces

 box will list all of the available communication interfaces. If the RFC1000 

is listed here, then the software has correctly recognized and is ready to use it.

Check that Windows recognizes the connected RFC1000 wireless transceiver.

If the software does not recognize the RFC1000, there may be a problem with Windows or the USB drivers.

1. 

In Windows, click 

Start

, right-click 

Computer

 and choose 

Properties.

2. 

Select 

Device Manager

 in the left hand column.

3. 

Double-click on 

Universal Serial Bus Controllers

4. 

Look for an entry for

 Data Logger Interface

.

5. 

If the entry is present, and there are no warning messages or icons, then windows has correctly  
recognized the connected RFC1000. 

6. 

If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need to 
be installed. USB drivers can be downloaded from the MadgeTech website.

Ensure that the USB end of the RFC1000 is securely connected to the computer.

1. 

If the cable is connected to the PC, unplug it and wait ten seconds. 

2. 

Reconnect the cable to the PC.

3. 

Check to make sure that the red LED is lit, indicating a successful connection.

Summary of Contents for Therm-A-lert

Page 1: ...Therm A lert Wireless Temperature Data Logger Product User Guide...

Page 2: ...s 3 Product Overview 4 Software Installation 5 Activating Deploying the Data Logger 6 Channel Programming 7 Product Maintenance 8 RMA Instructions 8 Troubleshooting 9 Compliance Information 10 General...

Page 3: ...are All active MadgeTech data loggers that are within range will automatically appear in the Connected Devices window 5 Select the data logger within the Connected Devices window and click the Claim i...

Page 4: ...er RFC1000 as well as from an RFC1000 to a Therm A lert Obstacles that interfere with or decrease the wireless signal could include but are not limited to building structures such as walls and metal b...

Page 5: ...rs to remotely monitor and manage groups of data loggers throughout a large facility or multiple locations from any internet enabled device Transmit real time data to the MadgeTech Cloud Services plat...

Page 6: ...position 0 is Off 1 is On 4 On the Windows PC launch the MadgeTech 4 Software 5 All active data loggers will be listed in the Device tab within the Connected Devices panel showing that the data logger...

Page 7: ...f the device 3 Change the DIP switches to match the desired channel 4 Return the device wireless mode to ON Channel Programming To configuring the channel settings of the RFC1000 wireless transceiver...

Page 8: ...on on the MadgeTech website 1 Go to www madgetech com rma 2 Sign into your account For new customers select Create an Account 3 Once signed in click on the Make New RMA button 4 Complete the applicabl...

Page 9: ...ess transceiver problems Check that the software properly recognizes the connected RFC1000 wireless transceiver If the wireless data logger is not appearing in the Connected Devices list it may be tha...

Page 10: ...conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l...

Page 11: ...in 55 mm x 133 mm x 33 mm Weight 3 56 oz 101 g Enclosure Material ABS Plastic Approvals US FCC CA IC CE General Specifications Range 20 C to 80 C 4 F to 176 F Resolution 0 01 C 0 018 F Calibrated Acc...

Page 12: ...DOC 1224036 00 REV 7 2016 04 26...

Reviews: