User’s Manual for Xplore
Version 2.53 - Version date: 15.02.2022
Page
5
of
46
Precautions for use
No studies have reported negative effects or side effects from using Xplore.
Risk analysis shows that users with one or more of the following conditions should consult their
doctor or therapist before starting to use the product to evaluate if benefits of use outweigh
possible risks of injury or negative side effects:
•
Major deformities (spine and lower limbs)
•
Severe or fixed contractures in the lower limbs (Hip and knee flexion contractures >40°, pes
equinus >25°)
•
Osteoporosis with previous or suspected spontaneous fractures of the lower extremities
•
Joint instability in the lower limbs (hips, knees, and ankles)
•
Circulatory disorders
•
Respiratory disorders
•
Cardiac disorders
•
Epilepsy with uncontrollable grand mal fits
•
Severe spasticity interfering with positioning and movement in the product
•
Pain while in standing weight-bearing position or when moving lower limbs
Training and Assistive Persons
Before the product is used, the supervisor must have received training (in accordance with Appendix
1) from an Made for Movement representative to monitor use and provide the user with any
necessary assistance. The representative is usually a Made for Movement employee or a distributor /
doctor / therapist who is appropriately trained and officially approved to work with the product.
The training includes a demonstration of all key features, transfer, attachment, safety, use and
troubleshooting as described in Appendix 1. In addition, the instructions for use must be read,
understood, used and retained for future reference.
Important:
The settings of the product must not be changed in any other way than
described in these operating instructions. Furthermore, the product may only be adjusted by an
Made for Movement representative who have been trained in the adjustment of the product.