MacroMedics ®
1. Warnings
• Inspect the device thoroughly before use. Do not use the device if damaged or if missing parts.
• Check individual components on MR-safety when used together in a patient positioning system.
• Do not use MR unsafe items in an MR environment.
•
Place and fix the device properly before positioning the patient.
• Rulers on the devices are only intended for reference of the device positioning of accessories. Rulers
shall not replace patient position verification.
• Handle with care to prevent damage.
• Ensure that the baseplate is correctly secured to the couch prior to use.
• In the USA, federal law restricts this device to sale by or on the order of a licensed healthcare
practitioner.
2. MRI Safety Information
2.1
MR UNSAFE
The following devices are MR unsafe:
• 123110 EAMIS
TM
Lite
2.2
MR SAFE
The following devices are MR safe:
• 123860 EAMIS
TM
Lite, MR
• 123400 SBRT Secure Arch, size L
• 123500 SBRT Secure Arch, size M
• 123600 SBRT Secure Arch, size S
• 123700 SBRT Secure Arch, size XS
2.3
MR CONDITIONAL
The following devices are MR conditional:
• All Vacuum Cushions distributed by MacroMedics are MR Conditional.
A patient with this device can be safely scanned in an MR system, meeting the following conditions:
•
A Static Magnetic field of 1.5 Tesla and 3 Tesla, with
•
Maximum spatial field gradient of 2,500 G/cm (25T/m)
•
Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2 W/Kg
(Normal operating mode)
Due to size and location of the metallic parts, no clinically significant RF heating is expected.