32
NANOCOLOR
®
FP-200 V 1.00 / 11.17
10.5 Executing a stored program
Call the desired test using its test number or from the test selection list. When the sipper
module is activated, you will see the
icon in the status bar of the photometer. In case
methods are programmed in the list of sipper methods (section 10.3), the settings stored with
the method will be automatically called used.
Note: Please consider the information in the instructions for use of the respective
test kit.
Hold the intake hose in the zero solution and press the
icon. The zero solution is pumped
into the flow cuvette and the zero measurement is taken automatically after the settling time
has expired. The time remaining until the measurement will be taken is shown in the display.
You can cancel the measurement by pressing
.
Hold the intake hose in the sample solution to be analyzed and press the
icon. The
sample solution is pumped into the flow cuvette and the sample measurement is taken
automatically after the settling time has expired. The time remaining until the measurement
will be taken is shown in the display. You can cancel the measurement by pressing
The flow cuvette is rinsed using distilled water or the next sample solution. The rinsing oper-
ation is initiated in the corresponding measurement menu by pressing
. The result of the
measurement is displayed in the result window.
You can enter various information on the sample using the icons below the result window.
The information entered is saved when you exit the menu or when the next measurement
is taken.