3.3 Scheme of operation for NANOCOLOR
®
tests
3.3.1 Scheme of operation when using tube tests with barcode
desired parameter: e.g. COD 160 test 0-26
1. Key
ESC
(if other parameters have been measured before)
17.11.11 16:06
>>
cuvette
>>
Method: ___
2. Insert coded test tube into the photometer.
Display of measuring range
17.11.11 16:06
0261 COD 160
15-160 mg/L O2
3. The interference filter is automatically turned into measuring position and the
cuvette is measured immediately.
17.11.11 16:06
0261 COD 160
0001 place
22 mg/L
O2
1+0
4. If desired, place other sample test tubes in photometer and measure; samples
are numbered in sequence.
3.3.2 Scheme of operation when using tube tests with or without barcode or for
manual operation
desired parameter: e.g. Ammonium 10 Test 0-04
1. Key
ESC
(if other parameters have been measured before)
17.11.11 16:06
>>
cuvette
>>
Method: ___
2. Insert coded test tube into the photometer or enter test number, keys
* * $
17.11.11 16:06
004_ AMMONIUM 10