Clinical Measurement Accuracy and Safety Verification
The product has passed clinical trials. The measured results of the infrared
forehead thermometer was compared with the measured results of mercury
thermometers, the deviation average
∆
t = 0.011°C not exceeding 0.3°C; the
clinical repeatability of the infrared forehead thermometer SR =0.100°C, not
exceeding 0.3°C. The measured results up to the laboratory standard and the
clinical standard.Therefore, the deviation average and the clinical repeatabili-
ty of the infrared forehead thermometer are complied with the regulatory re-
quirement ISO 80601-2-56. The conclusions are drawn from the clinical trials,
the accuracy and safety are compliedwith the regulatory requirement.
Warranty Information
• For repair under this warranty. Our authorized service agent must be advi-
sed of the fault within the period of the warranty. This warranty only covers
parts and labor service under normal operations. Any defect resulting from
natural causes, eg. flood, hurricane etc, is not covered in this guarantee.
This guarantee also does not cover damage incurred by use of the unit notin
accordance with the instructions, accidental damage, or being tampered or
serviced by unauthorized service agents.
• The fllowing will be excluded from this warranty-if the thermometer has
been misused, abused, or neglect in following the manual's instructions on
purpose and unauthorized repair or modifications.
• The device requires no calibration.
• The device is not repairable and contains no user serviceable parts.