background image

4. Position and Mount

Positioning and mounting EchoFlo is critical to the sensor performing 

correctly. EchoFlo should always be mounted perpendicular to the 

liquid in the pipe or tank, clear of any obstructions in the beam path. 

Not doing so will result in poor performance. 

Mounting Guide 

IMPORTANT:

1. Do not mount at an angle

2. Liquid should NEVER enter the  deadband

3. Mount at least 2” from the side wall

4. When pipe mounting, always install such that the bottom of the 

pipe is within the sensor’s range

5. Do not mount where obstacles will intrude on 2” beam width 

6. Do not mount in a vacuum

7. Never screw directly into the pipe or tank, ALWAYS use a  

NON-METALLIC fitting 

8. Always use an adapter with the minimal height possible

4.

 After you have entered configurations and tank values, click 

“Write to Unit”

 and send the configuration to EchoFlo. Now use  

EchoCal’s file management features to save your configuration by 

clicking 

“Save Config File”.

 A wiring diagram may be displayed by 

clicking 

“Wiring Diagram”.

NOTE: Be sure to record the “Sensor Height” and “Fill Height” 

before exiting EchoCal as these will be needed during the 

calibration process. 

3. Configuring EchoFlo

1.

 Start from the top of the “Config” area of 

EchoCal

 and work to the 

bottom, choosing the selections that are applicable.

NOTE: If the sensor is already configured 

with a Frequency output and it needs to 

be changed to a Voltage output (or vice 

versa) click “Clear Screen”. 

2.

 All configuration settings must be selected 

before you can continue. The recommended 

settings for connection to a MACE Series 3 

device are shown on the right. 

Always use 

supplied Viton® 

Gasket

Always use correct 

adapter (refer to 

Mounting Guide)

Always mount 

perpendicular 

to the liquid in 

the pipe or tank 

and clear of any 

obstructions in 

the beam path. 

Sensor Height:

Distance from the bottom of the pipe 

or tank to the bottom of the sensor. 

Units:

Units of measure

Fill Height:

Tank - Distance from the bottom 

of the tank to the maximum 

liquid height. 
Pipe - Distance of the  inner pipe 

diameter as shown below.

3.

 In the 

“Tank Levels” 

area of EchoCal select the: 

Fill Height (Pipe diameter)

Riser Distance (Greater than Sensor Deadband)

NOTE: The “Riser Distance” in a pipe MUST always be greater 

than the “Sensor Deadband” (refer to Table 1 over page). 

Reviews: