7. Calibrating an EchoFlo “Depth”
channel using FloCom
+
After following the previous configuration steps the EchoFlo must
be calibrated. Table 1 below lists relevant information for both
EchoFlo models including the
“Slope”
and
“Offset”
values for sensors
configured over their
“Effective Sensor Range”
.
TABLE 1: EchoFlo Slope and Offset Values
Model/Part# 1.25m (850-341)
2.5m (850-343)
Metric
US
Metric
US
Sensor Range
1.25m 49.2”
2.5m 98.4”
Deadband
0.05m
2”
0.1m
4”
Effective Sensor Range
1.2m 47.2”
2.4m 94.4”
Slope
0.00117 0.04609
0.00234 0.09218
Offset
-1.13766 -44.9875
-2.28 -89.975
Full Scale output
2000hz 2000hz
2000hz 2000hz
Zero output
976hz 976hz
976hz 976hz
3.
The
“Channel name”
dialogue
box will appear. Enter a channel
“Name”
in to the text field
(16 character maximum). This
“name”
will also be displayed on
the Series 3 device main LCD if
enabled. Click
“Continue”.
2.
Select
“Configure Channels”
and click
“Add”
and the
“New Channel”
dialogue
box will appear. Select the
“Depth”
channel type and click
“Continue”.
4.
In the
“New channel”
dialogue
box highlight the
“I/O Module”
option from which the named
channel will receive its value.
Click
“Continue”
and a new
dialogue box will appear with
the available input types listed.
5.
Select the
“Frequency”
input
type from the
“I/O Module @ 2”
list. Click
“Continue”
and then
select the appropriate input
from which the named channel
will receive its value.
6.
In this example the channel
named
“EchoFlo Depth”
will
be calculated using a sensor
attached to
“I/O Module @ 2”
on
“Digital input #1”
. Click
“Continue”
to complete the new
channel.
6.
The
“Configure channels”
dialogue box will re-appear
with the new channel listed.
6. Configuring an EchoFlo “Depth”
channel using FloCom
+
The EchoFlo is connected to a MACE Series 3 device by plugging the
pre-wired terminal strip into an I/O card within the device.
1.
Using MACE FloCom
+
software, connect to the device using the
serial port and configure the following settings:
• Enable
“Sensor power”
and select at least a 20 second warm up
time (a longer period may be required to record a stable reading
in some applications).
• Add an I/O module
NOTE: Refer to the relevant MACE Series 3 product manual for
detailed instruction on the above steps