Fig. 9: Remove Access Panel
Fig. 10: Remove HEPA Filter
Fig. 11: HEPA Filter (part #H0213)
Instruction Manual
MAN-040
12
www.macmedical.com
Replacing the HEPA Filter
1. Turn off the fan or unplug the cabinet for
power vented models.
2.
To access the HEPA filter (part #H0213),
remove the two screws (shown in green)
from the Filter Access Panel (Fig. 9 -
shown in yellow).
3. Remove the Filter Access Panel assembly
from the air vent panel (Fig. 10).
4.
Remove the HEPA filter from the
recessed filter box (Fig. 10).
5.
Insert a new HEPA filter into the recessed
filter box. Make sure the filter backing is
toward the fan.
6. Align the screw holes of the Filter Access
Panel assembly with the holes on the
bottom of the air vent panel.
7. Reinsert the screws into the Filter Access
Panel to secure.
8. Turn on the cabinet fan for power
vented models.